Remote Commands
R&S
®
FSW
958
User Manual 1173.9411.02 ─ 43
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
LIST:RANG2:INP:ATT:AUTO ON
Manual operation:
See
[SENSe:]LIST:RANGe<ri>:INPut:GAIN:STATe
<State>
This command turns the preamplifier for a spurious emission measurement range on
and off.
[SENSe:]LIST:RANGe<ri>:INPut:GAIN[:VALue]
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
LIST:RANG2:INP:GAIN:STAT ON
Manual operation:
See
[SENSe:]LIST:RANGe<ri>:INPut:GAIN[:VALue]
<Gain>
This command selects the preamplification level for the range.
The command requires option R&S
FSW-B24.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<Gain>
15 dB | 30 dB
The availability of preamplification levels depends on the
R&S
FSW model.
•
R&S
FSW8/13: 15dB and 30 dB
•
R&S
FSW26 or higher: 30 dB
All other values are rounded to the nearest of these two.
*RST:
OFF
Example:
LIST:RANG2:INP:GAIN 15
Configuring and Performing Measurements