Remote Commands
R&S
®
FSW
1147
User Manual 1173.9411.02 ─ 43
FORMat:DIMPort:TRACes
<Selection>
This command selects the data to be included in a data import file (see
For details on importing data see
Chapter 9.6.3, "How to Import Traces"
Parameters:
<Selection>
SINGle | ALL
SINGle
Only a single trace is selected for import, namely the one speci-
fied by the
on page 1147 com-
mand.
ALL
Imports several traces at once, overwriting the existing trace
data for any active trace in the result display with the same trace
number. Data from the import file for currently not active traces
is not imported.
The <trace> parameter for the
on page 1147 command is ignored.
*RST:
SINGle
Manual operation:
See
"Import All Traces/Import to Trace"
See
MMEMory:LOAD<n>:TRACe
<Trace>, <FileName>
This command imports trace data from the specified window to an ASCII file.
Suffix:
<n>
.
Parameters:
<Trace>
Number of the trace to be stored
(This parameter is ignored for
).
<FileName>
String containing the path and name of the import file.
MMEMory:STORe<n>:TRACe
<Trace>, <FileName>
This command exports trace data from the specified window to an ASCII file.
For details on the file format see
Chapter 9.6.6, "Reference: ASCII File Export Format"
Secure User Mode
In secure user mode, settings that are stored on the instrument are stored to volatile
memory, which is restricted to 256
MB. Thus, a "memory limit reached" error can occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
Analyzing Measurements (Basics)