Configuration
R&S
®
FSW-K7
84
User Manual 1173.9240.02 ─ 23
Setting the Reference Level from the Measurement (Meas->Ref)
................................86
Average Count (Number of Readings)
..........................................................................86
Using the power sensor as an external trigger
..............................................................86
.....................................................................................87
................................................................................................ 87
................................................................................................ 87
State
Switches the power measurement for all power sensors on or off. Note that in addition
to this general setting, each power sensor can be activated or deactivated individually
by the
setting on each tab. However, the general setting overrides the individual
settings.
Remote command:
Continuous Value Update
If activated, the power sensor data is updated continuously during a sweep with a long
sweep time, and even after a single sweep has completed.
This function cannot be activated for individual sensors.
If the power sensor is being used as a trigger (see
on page 86), continuous update is not possible; this setting is
ignored.
Remote command:
[SENSe:]PMETer<p>:UPDate[:STATe]
Select
Selects the individual power sensor for usage if power measurement is generally acti-
vated (
The detected
serial numbers
of the power sensors connected to the instrument are
provided in a selection list. For each of the four available power sensor indexes
("Power Sensor 1"..."Power Sensor 4"), which correspond to the tabs in the configura-
tion dialog, one of the detected serial numbers can be assigned. The physical sensor is
thus assigned to the configuration setting for the selected power sensor index.
By default, serial numbers not yet assigned are automatically assigned to the next free
power sensor index for which "Auto Assignment" is selected.
Alternatively, you can assign the sensors manually by deactivating the "Auto" option
and selecting a serial number from the list.
Remote command:
SYSTem:COMMunicate:RDEVice:PMETer<p>:DEFine
Input and Frontend Settings