Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
225
User Manual 1173.9240.02 ─ 23
Example:
PMET2:MTIM:AVER ON
Activates manual averaging.
PMET2:MTIM:AVER:COUN 8
Sets the number of readings to 8.
Manual operation:
See
"Average Count (Number of Readings)"
[SENSe:]PMETer<p>:MTIMe:AVERage[:STATe]
<State>
This command turns averaging for power sensor measurements on and off.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
PMET2:MTIM:AVER ON
Activates manual averaging.
Manual operation:
See
[SENSe:]PMETer<p>:ROFFset[:STATe]
<State>
This command includes or excludes the reference level offset of the analyzer for power
sensor measurements.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<State>
ON | 1
Includes the reference level offset in the results.
OFF | 0
Ignores the reference level offset.
*RST:
1
Example:
PMET2:ROFF OFF
Takes no offset into account for the measured power.
Manual operation:
See
[SENSe:]PMETer<p>[:STATe]
<State>
This command turns a power sensor on and off.
Suffix:
<p>
.
1...4
Power sensor index
Configuring the Measurement