Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
364
User Manual 1173.9240.02 ─ 23
11.8.2.2
Managing Limit Lines
Useful commands for managing limit lines described in the R&S
FSW User Manual:
●
MMEM:SEL[:ITEM]:LIN:ALL
●
MMEM:STOR:TYPE
●
MMEM:LOAD:TYPE
Remote commands exclusive to managing limit lines:
.................................................................................... 364
........................................................................................364
...................................................................................... 365
.......................................................................................365
CALCulate<n>:LIMit<k>:TRACe<t>:CHECk
......................................................................365
.............................................................................................. 366
............................................................................................ 366
CALCulate<n>:LIMit<k>:ACTive?
This command queries the names of
all
active limit lines.
Suffix:
<n>, <k>
.
irrelevant
Return values:
<LimitLines>
String containing the names of all active limit lines in alphabeti-
cal order.
Example:
CALC:LIM:ACT?
Queries the names of all active limit lines.
Usage:
Query only
Manual operation:
See
CALCulate<n>:LIMit<k>:COPY
<Line>
This command copies a limit line.
Suffix:
<n>
.
<k>
Parameters:
<Line>
1 to 8
number of the new limit line
<name>
String containing the name of the limit line.
Analyzing Results