<Type>
Detected power sensor type, e.g. "NRP-Z81".
<Interface>
Interface the power sensor is connected to; always "USB"
<SerialNo>
Serial number of the power sensor assigned to the specified
index
Example:
SYST:COMM:RDEV:PMET2:DEF '','NRP-Z81','',
'123456'
Assigns the power sensor with the serial number '123456' to the
configuration "Power Sensor 2".
SYST:COMM:RDEV:PMET2:DEF?
Queries the sensor assigned to "Power Sensor 2".
Result:
'','NRP-Z81','USB','123456'
The NRP-Z81 power sensor with the serial number '123456' is
assigned to the "Power Sensor 2".
Manual operation:
See
Configuring Power Sensor Measurements
CALibration:PMETer<p>:ZERO:AUTO ONCE
...................................................................220
CALCulate<n>:PMETer<p>:RELative[:MAGNitude]
........................................................... 221
CALCulate<n>:PMETer<p>:RELative[:MAGNitude]:AUTO ONCE
....................................... 221
CALCulate<n>:PMETer<p>:RELative:STATe
....................................................................222
[SENSe:]PMETer<p>:DCYCle[:STATe]
............................................................................ 222
[SENSe:]PMETer<p>:DCYCle:VALue
.............................................................................. 223
...................................................................................223
[SENSe:]PMETer<p>:FREQuency:LINK
...........................................................................223
.......................................................................................... 224
[SENSe:]PMETer<p>:MTIMe:AVERage:COUNt
................................................................ 224
[SENSe:]PMETer<p>:MTIMe:AVERage[:STATe]
...............................................................225
[SENSe:]PMETer<p>:ROFFset[:STATe]
...........................................................................225
.........................................................................................225
[SENSe:]PMETer<p>:UPDate[:STATe]
............................................................................ 226
......................................................................................... 226
............................................................................... 226
CALibration:PMETer<p>:ZERO:AUTO ONCE
This command zeroes the power sensor.
Note that you have to disconnect the signals from the power sensor input before you
start to zero the power sensor. Otherwise, results are invalid.
Suffix:
<p>
.
1...4
Power sensor index
Configuring the Measurement