Configuration
R&S
®
FSW-K7
125
User Manual 1173.9240.02 ─ 23
In addition to common absolute demodulation, the R&S FSW Analog Demodulation
application also provides demodulation results relative to user-defined or measured
reference values in the Result Summary.
The settings for the Result Summary can be defined individually for the different modu-
lation types (FM, AM, PM). For each modulation, a separate tab is provided in the dia-
log box.
Detector
Detector type for demodulation results
"+ Peak"
Positive peak
"- Peak"
Negative peak
"+/- Peak"
Autopeak
"RMS"
Root mean square
Remote command:
The detector is specified by the
DETector<det>
suffix in
CONFigure:RELative:AM|FM|PM:DETector<det>...
commands.
Mode
Defines the mode with which the demodulation result is determined.
The modes are similar to those for the entire trace (see
"Clear Write"
Overwrite mode: the detector value is overwritten by each sweep.
This is the default setting.
"Max Hold"
The maximum value is determined over several sweeps and dis-
played. The R&S
FSW saves each result only if the new value is
greater than the previous one.
"Average"
The average result is determined over all sweeps.
Remote command:
CONFigure:ADEMod:RESults:AM:DETector<det>:MODE
CONFigure:ADEMod:RESults:FM:DETector<det>:MODE
CONFigure:ADEMod:RESults:PM:DETector<det>:MODE
State
Activates relative demodulation for the selected detector. If activated, the demodulated
result is set in relation to the
Remote command:
CONFigure:ADEMod:RESults:AM:DETector<det>:STATe
CONFigure:ADEMod:RESults:FM:DETector<det>:STATe
CONFigure:ADEMod:RESults:PM:DETector<det>:STATe
Demodulation