Configuration
R&S
®
FSV/A3000 I/Q
Analyzer
86
User Manual 1178.8536.02 ─ 01
Adjusting settings automatically during triggered measurements
When you select an auto adjust function, a measurement is performed to determine
the optimal settings. If you select an auto adjust function for a triggered measurement,
you are asked how the R&S
FSV/A should behave:
●
(default:) The measurement for adjustment waits for the next trigger
●
The measurement for adjustment is performed without waiting for a trigger.
The trigger source is temporarily set to "Free Run" . After the measurement is com-
pleted, the original trigger source is restored. The trigger level is adjusted as fol-
lows:
–
For IF Power and RF Power triggers:
Trigger Level = Reference Level - 15 dB
–
For Video trigger:
Trigger Level = 85 %
Remote command:
[SENSe:]ADJust:CONFigure:TRIGger
Adjusting all Determinable Settings Automatically ( Auto All )
...................................... 86
Adjusting the Center Frequency Automatically ( Auto Frequency )
.............................. 86
Setting the Reference Level Automatically ( Auto Level )
............................................. 87
Resetting the Automatic Measurement Time ( Meastime Auto )
...................................87
Changing the Automatic Measurement Time ( Meastime Manual )
.............................. 87
................................................................................................ 87
................................................................................................ 88
Adjusting all Determinable Settings Automatically ( Auto All )
Activates all automatic adjustment functions for the current measurement settings.
This includes:
●
●
Note:
Auto measurement. For some measurements, the
Settings Automatically ( Auto All )
function determines the required measurement
parameters automatically. In this case, the progress of the auto measurement is indica-
ted in a message box.
See the R&S
FSV3000/ FSVA3000 base unit user manual.
Remote command:
Adjusting the Center Frequency Automatically ( Auto Frequency )
The R&S
FSV/A adjusts the center frequency automatically.
The optimum center frequency is the frequency with the highest S/N ratio in the fre-
quency span. As this function uses the signal counter, it is intended for use with sinus-
oidal signals.
At the same time, the optimal reference level is also set (see
Level Automatically ( Auto Level )"
This function is not available for input from the optional Digital Baseband Interface.
Adjusting Settings Automatically