Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
200
User Manual 1178.8536.02 ─ 01
Example:
TRAC:IQ:WBAN:MBW 82 MHZ
TRAC:IQ:WBAN:MBW?
Result if R&S
FSV3-B200 is active:
200000000
Manual operation:
See
8.4.6
Adjusting Settings Automatically
The commands required to adjust settings automatically in a remote environment are
described here.
[SENSe:]ADJust:CONFigure:DURation
............................................................................ 200
[SENSe:]ADJust:CONFigure:DURation:MODE
..................................................................201
[SENSe:]ADJust:CONFigure:HYSTeresis:LOWer
.............................................................. 201
[SENSe:]ADJust:CONFigure:HYSTeresis:UPPer
............................................................... 201
[SENSe:]ADJust:CONFigure:TRIGger
..............................................................................202
..........................................................................................202
[SENSe:]ADJust:ALL
This command initiates a measurement to determine and set the ideal settings for the
current task automatically (only once for the current measurement).
This includes:
●
Reference level
Example:
ADJ:ALL
Manual operation:
See
" Adjusting all Determinable Settings Automatically ( Auto
[SENSe:]ADJust:CONFigure:DURation
<Duration>
In order to determine the ideal reference level, the R&S
FSV/A performs a measure-
ment on the current input data. This command defines the length of the measurement
if
[SENSe:]ADJust:CONFigure:DURation:MODE
MANual
.
Parameters:
<Duration>
Numeric value in seconds
Range:
0.001 to 16000.0
*RST:
0.001
Default unit: s
Example:
ADJ:CONF:DUR:MODE MAN
Selects manual definition of the measurement length.
ADJ:CONF:LEV:DUR 5ms
Length of the measurement is 5 ms.
Configuring I/Q Analyzer Measurements