Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
167
User Manual 1178.8536.02 ─ 01
Suffix:
<n>
.
irrelevant
<p>
Power sensor index
Parameters:
<Unit>
DB | PCT
*RST:
DB
Example:
UNIT:PMET:POW:RAT DB
Triggering with Power Sensors
[SENSe:]PMETer<p>:TRIGger:DTIMe
..............................................................................167
[SENSe:]PMETer<p>:TRIGger:HOLDoff
........................................................................... 167
[SENSe:]PMETer<p>:TRIGger:HYSTeresis
.......................................................................168
[SENSe:]PMETer<p>:TRIGger:LEVel
............................................................................... 168
[SENSe:]PMETer<p>:TRIGger:SLOPe
............................................................................. 168
[SENSe:]PMETer<p>:TRIGger[:STATe]
.............................................................................169
[SENSe:]PMETer<p>:TRIGger:DTIMe
<Time>
This command defines the time period that the input signal has to stay below the IF
power trigger level before the measurement starts.
Suffix:
<p>
.
Power sensor index
Parameters:
<Time>
Range:
0 s to 1 s
Increment: 100 ns
*RST:
100 µs
Default unit: S
Example:
PMET2:TRIG:DTIMe 0.001
[SENSe:]PMETer<p>:TRIGger:HOLDoff
<Holdoff>
This command defines the trigger holdoff for external power triggers.
Suffix:
<p>
.
Power sensor index
Parameters:
<Holdoff>
Time period that has to pass between the trigger event and the
start of the measurement, in case another trigger event occurs.
Range:
0 s to 1 s
Increment: 100 ns
*RST:
0 s
Default unit: S
Example:
PMET2:TRIG:HOLD 0.1
Sets the holdoff time of the trigger to 100 ms
Configuring I/Q Analyzer Measurements