Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
161
User Manual 1178.8536.02 ─ 01
Example:
SYST:COMM:RDEV:PMET2:DEF '','NRP-Z81','',
'123456'
Assigns the power sensor with the serial number '123456' to the
configuration "Power Sensor 2".
SYST:COMM:RDEV:PMET2:DEF?
Queries the sensor assigned to "Power Sensor 2".
Result:
'','NRP-Z81','USB','123456'
The NRP-Z81 power sensor with the serial number '123456' is
assigned to the "Power Sensor 2".
Configuring Power Sensor Measurements
CALibration:PMETer<p>:ZERO:AUTO ONCE
................................................................... 161
CALCulate<n>:PMETer<p>:RELative[:MAGNitude]
............................................................161
CALCulate<n>:PMETer<p>:RELative[:MAGNitude]:AUTO ONCE
....................................... 162
CALCulate<n>:PMETer<p>:RELative:STATe
.....................................................................162
[SENSe:]PMETer<p>:DCYCle[:STATe]
............................................................................. 163
[SENSe:]PMETer<p>:DCYCle:VALue
...............................................................................163
................................................................................... 163
[SENSe:]PMETer<p>:FREQuency:LINK
........................................................................... 164
...........................................................................................164
[SENSe:]PMETer<p>:MTIMe:AVERage:COUNt
.................................................................164
[SENSe:]PMETer<p>:MTIMe:AVERage[:STATe]
................................................................ 165
[SENSe:]PMETer<p>:ROFFset[:STATe]
............................................................................165
..........................................................................................166
[SENSe:]PMETer<p>:UPDate[:STATe]
..............................................................................166
..........................................................................................166
................................................................................ 166
CALibration:PMETer<p>:ZERO:AUTO ONCE
This command zeroes the power sensor.
Note that you have to disconnect the signals from the power sensor input before you
start to zero the power sensor. Otherwise, results are invalid.
Suffix:
<p>
.
Power sensor index
Example:
CAL:PMET2:ZERO:AUTO ONCE;*WAI
Starts zeroing the power sensor 2 and delays the execution of
further commands until zeroing is concluded.
Usage:
Event
CALCulate<n>:PMETer<p>:RELative[:MAGNitude]
<RefValue>
This command defines the reference value for relative measurements.
Configuring I/Q Analyzer Measurements