
Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
166
User Manual 1178.8536.02 ─ 01
[SENSe:]PMETer<p>[:STATe]
<State>
This command turns a power sensor on and off.
Suffix:
<p>
.
Power sensor index
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
PMET1 ON
Switches the power sensor measurements on.
[SENSe:]PMETer<p>:UPDate[:STATe]
<State>
This command turns continuous update of power sensor measurements on and off.
If on, the results are update even if a single sweep is complete.
Suffix:
<p>
.
Power sensor index
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
PMET1:UPD ON
The data from power sensor 1 is updated continuously.
UNIT<n>:PMETer<p>:POWer
<Unit>
This command selects the unit for absolute power sensor measurements.
Suffix:
<n>
.
irrelevant
<p>
Power sensor index
Parameters:
<Unit>
DBM | WATT | W
*RST:
DBM
Example:
UNIT:PMET:POW DBM
UNIT<n>:PMETer<p>:POWer:RATio
<Unit>
This command selects the unit for relative power sensor measurements.
Configuring I/Q Analyzer Measurements