Analog Demodulation Option R&S
FSV–K7
R&S
®
FSV-K7/K7S
57
Operating Manual 1176.7561.02 ─ 06
Sweep Points
Opens an edit dialog box to enter the number of measured values to be collected dur-
ing one sweep.
●
Entry via rotary knob:
–
In the range from 101 to 1001, the sweep points are increased or decreased in
steps of 100 points.
–
In the range from 1001 to 32001, the sweep points are increased or decreased
in steps of 1000 points.
●
Entry via keypad:
All values in the defined range can be set.
The default value is 691 sweep points.
Remote command:
Capture Mode
This command defines how the size of the the capture buffer is defined for analog
demodulation.
In Auto mode, the size of the capture buffer is identical to the defined
In Manual mode, the size of the capture buffer can be defined manually (
Remote command:
Capture Time
Defines the size of the capture buffer for analog demodulation if
"Manual".
The maximum capture buffer size is 7.5
MSamples; the capture time depends on the
defined DBW. For a DBW of 25
kHz, for example, the maximum capture time is 240
s.
Remote command:
Demod Offset
This command defines an offset of the defined measurement time from the beginning
of the capture buffer for analog demodulation if
is set to "Manual". All
results are recalculated automatically based on the new data.
Remote command:
[SENSe:]ADEMod:CAPTure:DOFFset
2.2.8
Softkeys of the Trace Menu – TRACE key (Analog Demodulation)
The [TRACE] key is used to configure the data acquisition for measurement and the
analysis of the measurement data.
Trace 1/Trace 2/Trace 3/Trace 4/Trace 5/Trace 6
Selects the active trace (1, 2, 3, 4, 5, 6) and opens the "Trace Mode" submenu for the
selected trace.
Softkeys of the Analog Demodulation option