Remote Commands
R&S
®
FSMR3000
608
User Manual 1179.0116.02 ─ 02
Example:
SYST:DISP:LANG 'JA'
Switches the language of the instrument to Japanese.
12.8.5
Configuring the Network and Remote Control
The following commands are required to configure a network or remote control for the
R&S
FSMR3.
Useful commands for configuring remote control described elsewhere:
●
Remote commands exclusive to configuring a network and remote control
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess
............................................................... 608
SYSTem:COMMunicate:GPIB[:SELF]:RTERminator
.......................................................... 608
...............................................................................................609
................................................................................................ 609
..............................................................................................610
............................................................................................. 610
............................................................................................. 610
............................................................................................ 611
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess
<Address>
This command sets the GPIB address of the R&S
FSMR3.
Parameters:
<Address>
Range:
0 to 30
*RST:
(no influence on this parameter, factory default 20)
Example:
SYST:COMM:GPIB:ADDR 18
Manual operation:
See
SYSTem:COMMunicate:GPIB[:SELF]:RTERminator
<Terminator>
This command selects the GPIB receive terminator.
Output of binary data from the instrument to the control computer does not require
such a terminator change.
Parameters:
<Terminator>
LFEOI | EOI
LFEOI
According to the standard, the terminator in ASCII is <LF>
and/or <EOI>.
Configuring the R&S
FSMR3