Remote Commands
R&S
®
FSMR3000
609
User Manual 1179.0116.02 ─ 02
EOI
For binary data transfers (e.g. trace data) from the control com-
puter to the instrument, the binary code used for <LF> might be
included in the binary data block, and therefore should not be
interpreted as a terminator in this particular case. This can be
avoided by using only the receive terminator
EOI
.
*RST:
LFEOI
Example:
SYST:COMM:GPIB:RTER EOI
Manual operation:
See
SYSTem:DISPlay:LOCK
<State>
Defines whether the "Display Update" function remains available in remote operation
or not.
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
The function remains available.
ON | 1
The function is not available and the display is not updated dur-
ing remote operation.
*RST:
0
SYSTem:DISPlay:UPDate
<State>
This command turns the display during remote operation on and off.
If on, the R&S
FSMR3 updates the diagrams, traces and display fields only.
The best performance is obtained if the display is off during remote control operation.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
SYST:DISP:UPD ON
Manual operation:
See
SYSTem:ERRor:DISPlay
<State>
This command the error display during remote operation on and off.
If activated, the R&S
FSMR3 displays a message box at the bottom of the screen that
contains the most recent type of error and the command that caused the error.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Configuring the R&S
FSMR3