Remote Commands
R&S
®
FSMR3000
651
User Manual 1179.0116.02 ─ 02
SYSTem:HPADditional
<State>
Allows the use of HP commands
in addition to
SCPI commands for R&S
Parameters:
<State>
ON | OFF | 1 | 0
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
0
SYSTem:LANGuage
<Language>
This command selects the system language.
For details see
Chapter 11.3, "GPIB Languages"
Parameters:
<Language>
"SCPI" | "8560E" | "8561E" | "8562E" | "8563E" | "8564E" |
"8565E" | "8566A" | "8566B" | "8568A" | "8568A_DC" | "8568B" |
"8568B_DC" | "8591E" | "8594E" | "71100C" | "71200C" |
"71209A" | "PSA89600" | "PSA" | "PXA" | "FSP" | "FSU" |
"FSQ" | "FSV" | "FSEA" | "FSEB" | "FSEM" | "FSEK"
*RST:
SCPI
Example:
SYST:LANG 'PSA'
Emulates the PSA.
Manual operation:
See
Note
: If you use
"PSA89600"
, you must switch to an HP language first before return-
ing to SCPI (in remote operation only). For the identical language "PSA", this inter-
mediate step is not necessary.
SYSTem:PSA:WIDeband
<State>
This command defines which option is returned when the
*OPT?
query is executed,
depending on the state of the wideband option.
It is only available for PSA89600 emulation.
Parameters:
<State>
ON | OFF | HIGH
OFF
The option is indicated as "B7J"
ON
The 40
MHz wideband is used.
The option is indicated as "B7J, 140".
Emulating Other Instruments' Commands