Spectrum Measurements
R&S
®
ESR
261
User Manual 1175.7068.02 ─ 12
Limit Check 1-4 ← Sweep List dialog box ← Sweep List ← Spectrum Emission
Mask
Sets the type of limit check for all ranges.
For details on limit checks see
"Working with Limit Lines in SEM Measurements"
For details on limit checks see the base unit description "Working with Lines in SEM".
The limit state affects the availability of all limit settings (
Depending on the number of active power classes (see "Power Class" dialog box), the
number of limits that can be set varies. Up to four limits are possible. The sweep list is
extended accordingly.
Remote command:
[SENSe:]ESPectrum:RANGe<range>:LIMit<source>:STATe
Abs Limit Start ← Sweep List dialog box ← Sweep List ← Spectrum Emission
Mask
Sets an absolute limit value at the start frequency of the range [dBm].
This parameter is only available if the limit check is set accordingly (see
Remote command:
[SENSe:]ESPectrum:RANGe<range>:LIMit<source>:ABSolute:STARt
Abs Limit Stop ← Sweep List dialog box ← Sweep List ← Spectrum Emission
Mask
Sets an absolute limit value at the stop frequency of the range [dBm].
This parameter is only available if the limit check is set accordingly (see
Remote command:
[SENSe:]ESPectrum:RANGe<range>:LIMit<source>:ABSolute:STOP
Rel Limit Start ← Sweep List dialog box ← Sweep List ← Spectrum Emission
Mask
Sets a relative limit value at the start frequency of the range [dBc].
This parameter is only available if the limit check is set accordingly (see
Remote command:
[SENSe:]ESPectrum:RANGe<range>:LIMit<source>:RELative:STARt
Rel Limit Stop ← Sweep List dialog box ← Sweep List ← Spectrum Emission
Mask
Sets a relative limit value at the stop frequency of the range [dBc].
Measurements