Sample Application
R&S
®
DVMS1
30
Getting Started 2113.7619.02 ─ 12
6
Sample Application
Follow the instructions in this chapter to perform your first measurement with the
R&S
DVMS.
For an expanded selection of measurement examples and background information,
refer to the user manual or the help system.
In this measurement, the R&S
DVMS is used for basic compliance checks of one sin-
gle MPEG2 transport stream (TS) signal.
Since each R&S
DVMS option module provides at least one baseband transport
stream input (TS IN), you can perform the described measurement with every
R&S
DVMS independently of the option module type.
6.1
Requirements
The DVB TS
‑
ASI input signal used in this measurement is generated by either a TS
generator (i.e. R&S
DVSG) or by any other DVB-compliant TS
‑
ASI source.
The R&S
DVMS is locally operated by using an LCD monitor (DVI input required, at
least 1024
x
768 pixels) and a USB mouse.
6.2
Setting Up the R&S
DVMS
The required connectors are on the rear panel of the R&S
DVMS. For details on the
Chapter 3, "Interfaces and Connectors"
1. Connect the TS ASI output of the TS generator to the TS IN baseband input (BNC
connector) of the R&S
DVMS option module.
2. Connect the LCD monitor to the DVI
‑
D monitor output of the R&S
DVMS.
3. Plug the USB connector of the USB mouse in one of the two USB interfaces.
4. Switch on the R&S
DVMS and wait until the TS analyzer monitoring application is
running.
(Green LED [INSTRUMENT READY] at the front panel of R&S
DVMS)
The application window showing the default start screen is displayed.
Setting Up the R&S
DVMS