CONFigure:WCDMa:MEAS<i>:MEValuation:LIST:COUNt
<Segments>
Defines the number of segments in the entire measurement interval, including active
and inactive segments.
Parameters:
<Segments>
Range:
1 to 120
*RST:
10
Example:
See
Firmware/Software:
V1.0.5.3
Options:
R&S CMW-KM012
CONFigure:WCDMa:MEAS<i>:MEValuation:LIST:EOFFset
<Offset>
Defines the evaluation offset. The specified number of slots at the beginning of each
segment is excluded from the evaluation.
Set the trigger delay to 0 when using an evaluation offset (see
Parameters:
<Offset>
Range:
0 slots to 1024 slots
*RST:
0 slots
Example:
See
Firmware/Software:
V2.0.11
CONFigure:WCDMa:MEAS<i>:MEValuation:LIST:SEGMent<no>:SETup
<SegmentLength>, <Level>, <Frequency>[, <Retrigger>]
Defines the length and analyzer settings of a selected segment. In general this com-
mand must be sent for all segments measured.
Suffix:
<no>
.
1..1000
The segment number must not exceed the total number of seg-
ments measured (see