Remote Control Commands
R&S
®
NRX
376
User Manual 1178.5566.02 ─ 07
13.10
Zeroing
Further Information:
●
........................................................................................... 376
..................................................................................376
CALibration<undef>:ALL:ZERO:AUTO
............................................................................. 377
CALibration<Sensor>:ZERO:FAST:AUTO
.........................................................................377
CALibration<undef>:ALL:ZERO:FAST:AUTO
.................................................................... 377
CALibration<Sensor>:ZERO
[<auto>]
Performs zeroing for the power sensor connected to selected port.
Turn off all test signals before zeroing. An active test signal during zeroing causes an
error.
While zero calibration is in progress, no queries or other setting commands are
allowed, since the command is synchronous. Any communication attempt can run into
a timeout. Use
to recognize the end of a zeroing procedure.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<auto>
ON | OFF | ONCE | LFR | UFR
ON
Return value if a calibration is in progress.
OFF
Return value if no calibration is in progress.
ONCE
Starts zeroing.
LFR | UFR
Starts zeroing in a lower frequency range (LRF) or upper fre-
quency range (UFR), thus reducing the required time by half.
Not accepted by all power sensors. Consult the manual of the
power sensor concerned.
CALibration<Sensor>:ZERO:AUTO
[<auto>]
Performs zeroing using the signal at the power sensor input.
Turn off all test signals before zeroing. An active test signal during zeroing causes an
error.
While zero calibration is in progress, no queries or other setting commands are
allowed, since the command is synchronous. Any communication attempt can run into
a timeout. Use
to recognize the end of a zeroing procedure.
Zeroing