Command reference
R&S
®
ZNA
1111
User Manual 1178.6462.02 ─ 20
<Ch1>, <Ch2>, ...
Can only and must be specified for MANual optimized channel
ordering. Defines the order in which the channels are measured.
To get the optimization to work, you have to sweep the channels
manually, and in the specified order:
INIT<Ch2>
, ...
Example:
*RST;
*RST
creates channel 1, trace 1, and diagram 1.
:INITiate1:CONTinuous OFF
enable single sweep mode
CALC2:PAR:SDEF 'Ch2Tr1', 'S11'
CALC3:PAR:SDEF 'Ch3Tr1', 'S11'
CALC4:PAR:SDEF 'Ch4Tr1', 'S11'
create additional channels 2, 3 and 4 with one trace each
DISP:WIND1:TRAC2:FEED 'CH2TR1'
DISP:WIND1:TRAC3:FEED 'CH3TR1'
DISP:WIND1:TRAC4:FEED 'CH4TR1'
assign these traces to diagram 1
:CONF:CHAN:MEAS:OPT AUTO; :INITiate1:IMMediate:
ALL
optimize the switching times for the AUTOmatic channel order
1,2,3,4, and restart the sweep in all channels
:CONF:CHAN:MEAS:OPT MAN, 2,3,1,4; :INIT2; :
INIT3; :INIT1; :INIT4
optimize the switching times for channel order 2,3,1,4 and
restart the sweeps in the same order
Usage:
Setting only
CONFigure:CHANnel<Ch>[:STATe]
<Boolean>
Creates or deletes channel no.
<Ch>
and selects it as the active channel.
defines the channel name.
A channel created using
CONFigure:CHANnel<Ch>[:STATe] ON
can be configured
but has no trace assigned so that no measurement can be initiated. Use
CALCulate<Ch>:PARameter:SDEFine
<TraceName>, <Result>
to create a
channel and a trace. In remote control, it is possible to remove all channels. This is in
contrast to manual control where at least one channel with one diagram area and one
trace must be available.
Suffix:
<Ch>
.
Number of the channel to be created or deleted.
Parameters:
<Boolean>
ON - create channel no.
<Ch>
. If the channel no.
<Ch>
exists
already, it is not modified but selected as the active channel.
OFF - delete channel no.
<Ch>
.
*RST:
ON for channel no. 1 (created on *RST), OFF for all
other channels.
SCPI command reference