GUI reference
R&S
®
ZNB/ZNBT
709
User Manual 1173.9163.02 ─ 62
Skew
Enables/disables the skew measurement between the active trace and the
The skew measurement can only be performed if the following conditions are met for
both the active trace and the reference trace:
●
the trace format is
●
is enabled
●
the "Low Pass Step" time domain transformation
stimulus grid to be harmonic)
If these conditions are not met, a message is displayed:
Instead of performing the required configuration changes manually, you may also use
the "Auto Setup" button of the corresponding info field to let the firmware do this for
you.
Remote command:
Reference Trace
Selects the reference trace for the skew measurement.
Note that the reference trace must be in the same channel as the active trace.
See
for additional conditions on both the active and the reference trace.
Remote command:
Skew Position
Defines the percentage of the step size that defines the position of the reference points
on the current and the reference trace. The "Skew" value is the delta between the X
coordinates of these reference points.
Applic softtool