Physiolaser
olympic 500
Version 1.4 June 2007
11
Laser probes
Single laser probes
Nr.: Wave
length
Output power
Laser type
Impulse duration
NOHD
*)
m
a <= 1,5
Beam
divergence
510
785 nm max. 50 mW
Continuous (CW)
0,10
0,140 rad
511
810 nm max. 500 mW
Continuous (CW)
0,29
0,157 rad
512
670 nm max. 200 mW
Continuous (CW)
0,14
0,209 rad
513
655 nm Max. 50 mW
Continuous (CW)
0,09
0,157 rad
515
904 nm max.90 W
Impulse 200 ns
0,05
0,190 rad
Multi-cluster probes
Nr.: Wave
length
Output power
Laser type
Impulse duration
NOHD
*)
m
a <= 1,5
Beam
divergence
516
904 nm 5 x max. 30 W
Impulse 200 ns
0,06
0,190 rad
517 785
nm
655 nm
655 nm
4 x max. 55 mW
4 x max. 40 mW
4 x max. 5 mW
Continuous (CW)
Continuous (CW)
Continuous (CW)
0,10
0,09
0,04
0,140 rad
0,157 rad
0,155 rad
518
904 nm 12 x max. 10 W Impulse 200 ns
0,05
0,190 rad
*) Safety distance (NOHD). Distance at which the radiation intensity or the radiation is equivalent to the limiting value
of the maximum permitted radiation of the cornea (MZB).
For more information, please refer to the device plate.
The controller of the
Physiolaser
olympic
will recognize automatically which probe is
connected and will adjust all relevant settings. Please observe that when the output is
set, the laser diodes’ maximum laser output may differ from one another. The probes
can be identified through their name plates, the first two digits in each number specify
the maximum laser output in mW (CW) or W (imp). You will find the recognized probe
shown in the display as: