2. Programming (cont.)

To add a device

 to RG43 memory, just trigger an alarm from the 

device you want to add. The indication LED for the device type will 
turn on to indicate the memory slot now contains this device type 
(Call, Pad, or Sensor). Multiple devices can be added to the same 
memory slot. Adding a wireless reset button to a memory slot will 
allow the button to reset of any active alarm in this slot.

To delete a device

 from RG43 memory, select the memory slot 

and hold RESET for ~5s. You should see the indication LED turn off 
to indicate the device is no longer present in memory.

2.3 Auto-Programming

The RG43 panel now supports auto-programming when used
together with an RG2 Companion Wireless Monitor. 
RG2 Companion Wireless Monitor must be set to t=2 or t=3 for 
this to work.

On the 

RG2

:

• Press P PAUSE for ~2s
• For t=2, use PAUSE to adjust tens digit and RESET to adjust the 
units if the monitor will be moved to a new room
• For t=3, the display settings cannot be adjusted, and the 

 

 

 

 

 

 

 

monitor will display [AP] to indicate it is ready for auto-pairing.
• Take the monitor next to the RG43 panel to synchronise the 
displays.

On 

RG43

:

• Hold P REVIEW for ~5s to enter pairing mode
• The RG43 will briefly display [AP], and all devices paired with the 
RG2 monitor will be uploaded to the panel.
• The panel will indicate the number of devices that are received 
from the RG2 monitor, followed by the new display ID or IDs.

Example 1. If an RG2 monitor with a bed and chair sensor programmed in and are set to t=2, then is uploaded with 
new display [34], the RG43 panel will first show [d2] to indicate that two devices are paired into this display, followed 
by [34] to confirm what will now display when a device is triggered.
Example 2. If an RG2 monitor set to t=3 with two bed sensors programmed to  displays [34] and [35], the 
RG43 will first show [d2] to indicate that two devices  are being paired, followed by [34], [35] to confirm 
what will now display when these devices are triggered.

Summary of Contents for RG43

Page 1: ...Display Monitor RG43 User Instructions ...

Page 2: ......

Page 3: ...ge LED display from a distance and the panel can also forward signals to a Caregiver Alert Pager that will display call type and device ID Alarms can be reset using a wireless reset button cordless monitor pager or at the panel Straightforward programming allows users to quickly set up the system and easily add remove devices based on changes in daily requirements ...

Page 4: ... pre configured to operate with your kit If not refer to programming section below for details 1 4 Adjust volume to desired level by using the wheel on the right side of the monitor Tracking should only be ON if mobile devices wheelchairs or pendant call buttons are in use Otherwise set this switch to OFF Switch on the button lock to prevent accidental activation of the buttons on the front of the...

Page 5: ...can be useful if several panels are in use and will ensure pagers only receive alarms from their area Tap SELECT1 until the tens digit displays A Use SELECT2 to set 0 7 This setting must match whatever has been selected on the wireless devices in the area Pause Setting This allows a caregiver to pause the panel for a programmable interval allowing patients to be removed from their beds without tri...

Page 6: ...t A Alarm delays 10min after bed exit B Alarm delays 15min after bed exit C Alarm delays 20min after bed exit D Alarm delays 25min after bed exit E Alarm delays 30min after bed exit F Alarm delays 35min after bed exit Pager Code Setting a pager code can be useful if several panels are in use and will ensure pagers only receive alarms from their area Tap SELECT1 until the tens digit displays C Use ...

Page 7: ...t and RESET to adjust the units if the monitor will be moved to a new room For t 3 the display settings cannot be adjusted and the monitor will display AP to indicate it is ready for auto pairing Take the monitor next to the RG43 panel to synchronise the displays On RG43 Hold PROGRAM REVIEW for 5s to enter pairing mode The RG43 will briefly display AP and all devices paired with the RG2 monitor wi...

Page 8: ...and a beep will sound if AC is lost Panel Battery This indicates the status of the panel only It will flash if there is low battery or batteries have not been inserted This LED can only be reset by changing the batteries The two LEDs next to the large digits indicate the status of wireless devices Signal Lost will turn on if tracking is active and the panel cannot detect one of the devices in memo...

Page 9: ...to a memory slot to give a totally wireless falls man agement solution If multiple devices are programmed into a single memory slot this will reset any alarm from the slot into which it has been paired When using wireless reset buttons Ramblegard recommends holding the button for 3s to ensure the panel receives the signal Cordless Monitors Cordless RG2 monitors can now be used as a kit together wi...

Page 10: ...r local distributor with details for further assistance Signal Lost If a device is displaying as signal lost on the RG43 panel you can silence this indication by pressing RESET The SIGNAL LOST LED will continue to flash until the device has been brought back into range or is removed from memory If you would like to remove this device from the panel memory hold REVIEW RESET for 3s The RG43 panel wi...

Page 11: ......

Page 12: ...T 028 9448 3971 www ramblegard com S E CareTrade t a Ramblegard Elmfield House Rathenraw Industrial Estate Greystone Road Antrim BT41 2SJ ...

Reviews: