2. Programming
2.1
Panel Setting
To access the panel settings menu, press and hold the PROGRAM
+ SELECT1 buttons for ~5s. The display will beep once. Use the
SELECT1 button to cycle between the settings, and SELECT2 to
change.
Call Forward
:
To configure the call-forward feature, tap SELECT1
until the tens digits displays “t”. Then use SELECT2 to adjust the
setting:
Set P = Pause Setting
0
OFF
1
Not Used
2
Not Used
3
Signals to RG41 (Pager Unit)
4
Panel receives from master, does not transmit
Site Code
:
Setting a site code can be useful if several panels are
in use, and will ensure pagers only receive alarms from their area.
• Tap SELECT1 until the tens digit displays “A”.
• Use SELECT2 to set 0-7
• This setting must match whatever has been selected on the
wireless devices in the area
Pause Setting
:
This allows a caregiver to pause the panel for a
programmable interval, allowing patients to be removed from
their beds without triggering an active alarm.
• Tap SELECT1 until the tens digit displays “P”.
• Use SELECT2 to change the setting:
Set P = Pause Setting
0
Panel cannot be paused
1-9
Pause 1-9min
A
Pause for up to 20min
B
Pause for up to 40min
C
Pause for up to 60min
D
Pause for up to 80min
E
Pause for up to 100min
F
Pause for up to 120min