PROGRAMMING
4-23
4.7.2. Enable / Disable Sensor
This function allows the user to selectively enable or disable
individual sensors in the QRAE PLUS monitor. When a sensor
is disabled, the unit will not measure or display gas
concentrations of that type.
1. “Enable / Disable Sensors?” is the second sub-menu item.
2. Pressing the
[Y/+]
key causes the QRAE PLUS to display
all the sensors installed in the monitor. A previously enabled
sensor is marked with an “*” next to the sensor name. The
cursor will blink at the first sensor location. Press the
[Y/+]
key to enable the sensor or the
[N/-]
key to disable the
sensor. A previously disabled sensor will show “----” instead
of the sensor name.
CO*
H
2
S*
LEL* pick
OXY*
3. Press the
[MODE]
key momentarily to move to the next
sensor location. Repeat Step 2 until all the appropriate
sensors are enabled.
4. Press and hold the
[MODE]
key for one second to save
your changes. The display will show “Save?” To confirm
the new selection, press the
[Y/+]
key. Press the
[N/-]
or
[MODE]
key to discard the change and move to the next
sub-menu.
Summary of Contents for QRAE PLUS PGM-2000
Page 5: ...NOTE THE SYMBOL MEANS CAUTION RISK OF DANGER CONSULT OPERATION AND MAINTENANCE MANUAL ...
Page 8: ......
Page 12: ...GENERAL INFORMATION 1 4 non condensing ...
Page 33: ...OPERATION 2 20 A new datalog event is created when the datalogging is resumed ...
Page 34: ......
Page 65: ......
Page 67: ......
Page 73: ......
Page 81: ...TROUBLESHOOTING 8 8 ...
Page 83: ...APPENDIX A A 2 ü Function allowed Password required ...
Page 87: ...APPENDIX B B 3 Xylene p 2 8 1 1 Lower Explosion Limit of the gas in volume percent ...