To enter
<
Customer Calibration>
,
press
and then
:
„
Parameters /
Customer Calibration”
.
21.1. Calibration procedure
•
Enter
submenu
<
Customer Calibration>
according to ch. 21
and select:
“
Platform 1 /
Calibration”
,
•
After entering the parameter the following message box appears:
•
Take the load off the pan of platform 1,
•
Press
button
. The following message appears during
adjusting start mass:
„Evaluation of starting mass”
,
•
After the procedure has been completed the following message
box appears:
•
Put calibration mass on pan of platform 1 and then select
,
57