- 68 -
19.4.11. Switch on continuous transmission in current unit
Syntax:
CU1 CR LF
Possible answers:
CU1_I CR LF
- command comprehended but cannot be executed
CU1_A CR LF
MASS FRAME
- command comprehended and in progress
- mass value in current unit is returned
Frame format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S
U
I
stability
space
sign
mass
space
unit
CR
LF
19.4.12. Switch off continuous transmission in current unit
Syntax:
CU0 CR LF
Possible answers:
CU0_I CR LF
- command comprehended but cannot be executed
CU0_A CR LF
- command comprehended and executed
19.4.13. Lock the scale keyboard
Syntax:
K1 CR LF
Possible answers:
K1_I CR LF
- command comprehended but cannot be executed
K1_OK CR LF
- command executed
Caution:
This command is not remembered after restart