Guidance and manufacturer’s declaration – electromagnetic immunity 

The  TH8xyz series  is intended for use in the electromagnetic environment specified below. The 
customer or the user of the  TH8xyz series  should assure that it is used in such an environment. 

Immunity test

 

IEC 60601 test 

level

 

Compliance 

level

 

Electromagnetic environment 

– 

guidance

 

Electrostatic     
discharge 
(ESD) 

IEC 61000-4-2 

6 kV contact 
8 kV air 

6 kV 
contact 
8 kV air 

Floors should be wood, concrete 
or ceramic tile. If floors are 
covered with synthetic material, 
the relative humidity should be at 
least 30 %. 

Electrical fast 
transient/burst 
 
IEC 61000-4-4 

2 kV for power 
supply lines 
1 kV for 
input/output lines 

Not applicable

Mains power quality should be that of 
a typical commercial or hospital 
environment. 

Surge 
IEC 61000-4-5 

1 kV line(s) to 
line(s) 
2 kV line(s) to earth 

Not applicable

Mains power quality should be that of 
a typical commercial or hospital 
environment. 

interruptions and 
voltage 
variations on 
power supply 
input lines 
 
IEC 61000-4-11 

<5 % UT 
(>95 % dip in UT) 
for 0,5 cycle 

 

40 % UT 
(60 % dip in UT) 
for 5 cycles 

 

70 % UT 
(30 % dip in UT) 
for 25 cycles 

 

<5 % UT 
(>95 % dip in UT) 
for 5 sec 

Not applicable

Mains power quality should be that of 
a typical commercial or hospital 
environment. If the user of the 
TH8xyz series   requires 

continued 

operation during power mains 
interruptions, it is recommended that 
the TH8xyz series be powered from 
an uninterruptible power supply or a 
battery. 

Power frequency 
(50/60 Hz) 
magnetic field 
 
IEC 61000-4-8 

3 A/m 

3 A/m 

Power frequency magnetic fields 
should be at levels characteristic of a 
typical location in a typical 
commercial or hospital environment. 

NOTE UT is the a.c. mains voltage prior to application of the test level. 

 
 

Recommended separation distances between 

portable and mobile RF communications equipment and the 

ME EQUIPMENT or ME SYSTEM

 

The  TH8xyz series  is intended for use in an electromagnetic environment in which radiated RF 
disturbances are controlled. The customer or the user of the  TH8xyz series  can help prevent 
electromagnetic interference by maintaining a minimum distance between portable and mobile RF 
communications equipment (transmitters) and the  TH8xyz series  as recommended below, 
according to the maximum output power of the communications equipment.

 

Rated maximum 

output power of 

transmitter 

Separation distance according to frequency of transmitter 

m

 

150 kHz to 80 MHz 

d = 

1,2 

 

80 MHz to 800 MHz

d = 

1,2 

 

800 MHz to 2,5 GHz 

d = 

2,3 

 

0,01

 

N/A 0,12 0,23 

0,1

 

N/A

 

0,38

 

0,73

 

1

 

N/A

 

1,2

 

2,3

 

10

 

N/A

 

3,8

 

7,3

 

100

 

N/A

 

12

 

23

 

For transmitters rated at a maximum output power not listed above, the recommended separation 
distance 

in meters (m) can be estimated using the equation applicable to the frequency of the 

transmitter, where 

is the maximum output power rating of the transmitter in watts (W) according 

to the transmitter manufacturer.   
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. 
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected 
by absorption and reflection from structures, objects and people.

 

 

 

 

 

 

 