Guidance and manufacturer’s declaration – electromagnetic immunity
The TH8xyz series is intended for use in the electromagnetic environment specified below. The
customer or the user of the TH8xyz series should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environment
–
guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV
contact
8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for power
supply lines
1 kV for
input/output lines
Not applicable
Mains power quality should be that of
a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
1 kV line(s) to
line(s)
2 kV line(s) to earth
Not applicable
Mains power quality should be that of
a typical commercial or hospital
environment.
interruptions and
voltage
variations on
power supply
input lines
IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
Not applicable
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the
TH8xyz series requires
continued
operation during power mains
interruptions, it is recommended that
the TH8xyz series be powered from
an uninterruptible power supply or a
battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
Recommended separation distances between
portable and mobile RF communications equipment and the
ME EQUIPMENT or ME SYSTEM
The TH8xyz series is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the TH8xyz series can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the TH8xyz series as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d =
1,2
80 MHz to 800 MHz
d =
1,2
800 MHz to 2,5 GHz
d =
2,3
0,01
N/A 0,12 0,23
0,1
N/A
0,38
0,73
1
N/A
1,2
2,3
10
N/A
3,8
7,3
100
N/A
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance
d
in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where
P
is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.