Recommended separation distances between 

portable and mobile RF communications equipment and the 

TH6xyz series

 

The 

TH6xyz series

  is intended for use in an electromagnetic environment (for home healthcare) in which radiated RF disturbances are controlled. The 

customer or the user of the 

TH6xyz series

  can help prevent electromagnetic interference by maintaining a minimum distance between portable and 

mobile RF communications equipment (transmitters) and the 

TH6xyz series

  as recommended below, according to the maximum output power of the 

communications equipment. 

Rated maximum output power of 

transmitter 

Separation distance according to frequency of transmitter 

150 kHz to 80 MHz 

d = 1,2 

 

80 MHz to 800 MHz 

d = 1,2 

 

800 MHz to 2,7 GHz 

d = 2,3 

 

0,01 N/A 0,12 0,23 

0,1 N/A 0,38 0,73 

1 N/A 1,2 2,3 

10 N/A 3,8 7,3 

100 N/A  12  23 

For transmitters rated at a maximum output power not listed above, the recommended separation distance 

in meters (m) can be estimated using the equation 

applicable to the frequency of the transmitter, where 

is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.   

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. 
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

 

 

Manufacturer’s declaration-electromagnetic immunity 

Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment 

The 

TH6xyz series

  is intended for use in the electromagnetic environment (for home healthcare) specified below.The customer or the user of the 

TH6xyz series

  should assure that it is used in such an environment. 

Test frequency 

(MHz) 

Band 

a)

 

(MHz) 

Service 

a)

 Modulation 

b)

 

Maximum power

(W) 

Distance 

(m) 

IMMUNITY TEST 

LEVEL 

(V/m) 

Compliance LEVEL

(V/m) 

(for home healthcare)

385 380–390  TETRA 

400 

Pulse 

modulation b) 

18 Hz 

1,8 0,3  27 

27 

450 430–470 

GMRS 460, 

FRS 460 

FM c) 

±5 kHz deviation

1 kHz sine 

2 0,3  28 

28 

710 

704–787 

LTE Band 13,17 

Pulse 

modulation b) 

217 Hz 

0,2 0,3 

745 

780 

810 

800–960 

GSM 800/900, 

TETRA 800, 

iDEN 820, 

CDMA 850, 

LTE Band 5 

Pulse 

modulation b) 

18 Hz 

2 0,3  28 

28 

870 

930 

1720 

1700– 
1990 

GSM 1800; 

CDMA 1900; 

GSM 1900;   

DECT; 

LTE Band 1, 3, 

4, 25; UMTS 

Pulse 

modulation b) 

217 Hz 

2 0,3  28 

28 

1845 

1970 

2450 

2400– 
2570 

Bluetooth,  

WLAN, 

802.11 b/g/n,   

RFID 2450,   

LTE Band 7 

Pulse 

modulation b) 

217 Hz 

2 0,3  28 

28 

5240 

5100– 
5800 

WLAN 802.11 

a/n 

Pulse 

modulation b) 

217 Hz 

0,2 0,3 

5500 

5785 

NOTE    If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME EQUIPMENT or ME
SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3. 

a) For some services, only the uplink frequencies are included. 
b) The carrier shall be modulated using a 50 % duty cycle square wave signal. 

c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not  represent actual modulation,

it would be worst case. 

 