4.1.1
Measuring body temperature on the forehead:
Press the “ON/MEM” button to power on the device.
Forehead mode is the default mode.
You can see the icon on the screen and hear two
beep sounds (see figure 1). In this mode, you can hold the thermometer within
≤
1 cm from the central forehead (Fig. 4) and press the "SCAN" button
to get the forehead measurement. The time consuming for measurement might be 1 second. After two short beeps are heard, the icon stops
flashing to be ready for next measurement.
** When you get a new temperature reading in forehead mode, the last reading will be shown on the screen (at top right corner) with the
icon.**
4.2 Measuring temperature under Surface mode:
4.2.1 When you press the “SCAN” button, you will get the real time temperature immediately. If you press and hold the “SCAN” button, the reading of measurement
will be continuous updated.
4.2.2 Applications include temperature measurements for Water, Milk, Cloth, Skin or other object.
* Note: This mode shows the actual, unadjusted surface temperatures, which is different from the body temperature.
4.3 Get the room temperature
4.3.1 For room temperature, the thermometer should be placed on the table and avoid the direct sunshine or air conditioner flow on.
4.3.2 When in power off, room temperature will always be shown on the screen with icon.
4.3.3 The room temperature can be reference after 15min later.
5. After Measurement:
5.1 Power off: Device will automatically shut off if left idle for more than 1 minute to extend battery life. The LCD will only show the room temperature.
5.2 Clean the probe after each use to ensure an accurate reading and avoid cross contamination. (See the section of Care and Cleaning for details.)
5.3 Remember to put on the Probe Cap when not in use.
Important Notes
Care and cleaning
The probe is the most delicate part of the thermometer. Use with care when cleaning the probe to avoid damage.
a. After each use, please use alcohol swabs or cotton swabs moistened with 70%~75% alcohol to clean the probe lens.
b. Allow the probe to fully dry for at least 1 minute.
c. Keep the unit dry and away from any liquids and direct sunlight.
d.
Storage temperature range: It should be stored at room temperature between -20~+50°C,
RH
≤
85%
e. The Probe should not be submerged into liquids.
Holding the thermometer too long may cause a higher ambient temperature reading of the probe.
This could make the body temperature measurement lower than usual.
Battery replacement
This device is supplied with one lithium cell CR2032x1
○
1
Open the battery cover: Insert a pointed object, such as a pen, into battery cover pick hole.
At the same time, use thumb to push battery cover out.
○
2
Hold the device and flip the battery out with a small screwdriver
○
3
Insert the new battery under the metal hook and press down until you hear a “click” sound.
○
4
Replace the battery cover
The positive (+) side up and the negative (-) side pointed down.
Trouble shooting:
Error Message
Problem
Solution
Error 5~9, the system is not functioning properly.
Unload the battery, wait for 1 minute and repower it. If the message
reappears, contact the retailer for service.
Measurement before device stabilization.
Wait
for “Er1” to disappear.
The ambient temperature is not within the range between 10
o
C and 40
o
C (50
o
F ~104
o
F).
Allow the thermometer to rest in a room for at least 15 minutes at
room temperature: 10
o
C and 40
o
C (50
o
F ~104
o
F).
(1)
In Forehead mode: Temperature taken is higher than +42.2
o
C (108°F)
(2)
In Surface mode: Temperature taken is higher than +80
o
C (176°F)
Please select the target within specifications. If a malfunction still
exists, please contact the nearest retailer.
(1)
In Forehead mode: Temperature taken is lower than +34
o
C (93.2°F)
(2)
In Surface mode: Temperature taken is lower than –22
o
C (-7.6°F)
Device cannot be powered on to the ready stage.
Change with a new battery.
Warranty:
Warranty: 12 months
Manufacture Date: as the serial number (please open the battery cover, it is shown on the inside of the device.)
Ex.SN:E512A000001, the first “E” is External, the second number “5” is the last number of manufacture year, the third and the fourth number “12” is the manufacture
month, the others is the serial number..
Note: The thermometer is calibrated at the time of manufacture. If you question calibration mode, the accuracy of temperature measurements or unexpected
events at any time, please contact the dealers or nearest service address.
Please report to the manufacturer and the competent authority of the Member State in which you are established about any serious incident that has
occurred in relation to this device.
Warning: No modification of this equipment is allowed.
Please read the instructions for use
BF type applied part
IMPORTER
Company
Name?
Company Name?
General Information: Address?
Website?
Radiant Innovation Inc. Http://www.radiantek.com.tw
Add: 1F, No.3, Industrial East 9
th
Road, Science-Based Industrial Park,
HsinChu, Taiwan 300.
Medical Technology Promedt Consulting GmbH
Add: Altenhofstrasse 80, D-66386 St. Ingbert,
Germany
Date of issue: 2020-09-16
Ref No.: 092020
Figure 4