Room Temperature

 

Suitable ambient temperature is important for the baby and patient, the thermometer always help you recognize the 
room temperature. 

 When in power off, room temperature will always be shown on the screen with   icon. 

 Please see the “Use of the thermometer” section to learn how to get the correct room temperature.   

 

Forehead 
Temperature 

The thermometer has been designed for home use. It’s not meant to replace a visit to the doctor. Please also remember 
to compare the measurement result to your regular body temperature. 

 Please see the “Use of the thermometer” section to learn how to measure the body temperature. 

 

Scan Mode 

The Scan mode shows the actual, unadjusted surface temperatures, which is different from the body temperature. It can 
help you to monitor if the object temperature is suitable for the baby or patient, for example the baby’s milk. 

 Please see the “Use of the thermometer” section to to learn how to measure the object temperature. 

 

Fever Indication

 

If thermometer detects a body temperature 

 37.5°C under forehead mode, there will be a long beep sound followed by 

three short beep sound to warn the user for potential fever. 

 

Last Reading 

When you get a new temperature reading in forehead mode, the last reading will be shown on the screen (at top right 
corner) with the 

 icon. 

Memory Locations

 

There are total 25 sets memories for body temperature.   

 When in power on, press the “ON/MEM” button to see the temperature stored with 

 icon.  

 

°C / °F Switch

 

 In “power off” mode (The LCD only shows the room temperature). Press and hold the “SCAN” button, then 

immediately press and release the “ON/MEM” button. After 3 seconds, icon “°C” will be switched to icon “°F”. (You may 
use the same way to switch back to “°C”) 

 

 

 

 

Use of the thermometer

 

Note: If there is any temperature difference between the places where the device is stored and where you are going to   
          measure, subject and the device should stay in the same room for at least 15 minutes before measurement. 

1. Always make sure the probe is clean, and without damage.   
2. Power on: Press “ON/MEM” button.(see figure 1)   

3. Mode selection:   

3.1

 

Forehead mode

: Press the “ON/MEM” button, the default mode of thermometer is forehead mode. The thermometer is   

ready for use after you see the    icon on the screen and hear two beep sounds (see figure 1). In   
this mode, you can get the body temperature by forehead measurement. 

 

3.2 Scan Mode

After power on, press and hold the “ON/MEM” button, and press the “SCAN” button   

one time for “Infrared thermometer” mode

 to see

 

  icon on your LCD display (see figure2).   

In this mode, you can get the target surface temperature.

 

4. Temperature taking:   

4.1 Forehead temperature measuring, points for attention:

   

Remarks: a. Forehead temperature is displayed in oral mode. This mode converts the forehead temperature to display its “oral-equivalent” Value. 

b. Before the measurement, the subject should stay in a stable environment for 5mins and avoid exercise, bath for 30mins.   
c. Remember to keep the forehead area clean and away from sweat, cosmetics and scar while taking temperature. 
d. The “Clinical Bias” is -1.3 ~ -1.5°C. 

e. The “Limits of Agreement” is 0.31. 
f. The "Repeatability" is 0.17°C. 

TH60F Non-contact Forehead Thermometer 

Intended use: The non-contact forehead thermometer is electronic thermometer using an infrared detector 

(thermopile detector) to detect body temperature from forehead center in people of all ages. 

Intended operator: Have eight years of education, no maximum.

 

 
 
 
 

 

 

Specifications 

 

Temperature measurement range:   

Forehead mode: 34~42.2°C (93.2~108°F), Surface mode: -22~80°C (-7.6~176°F) 

 

Operating temperature range: 10~40°C (50~104°F), 15%~85% RH 

 

Storage temperature range: It should be stored at room temperature between -20~+50°C, RH

85% 

Transportation temperature shall be less than 70°C, RH

95% 

 

Atmospheric pressure: 800~1013 hPa 

 

Comply with ASTM E1965-98, EN ISO 80601-2-56, IEC/EN60601-1-2(EMC), IEC/EN60601-1(Safety) 
standards, ISO10993, RoHS   

 

Accuracy: 
For forehead mode:

 

±0.2°C (0.4°F) within 35~42°C (95~107.6°F), ±0.3°C (0.5°F) for other range. 

For surface mode: 

±

0.3°C (0.5°F) within 22~42.2°C (71.6~108°F), others 

±

4% or 

±

2°C (4°F)   

              whichever is greater. 

 

Fever indication & Temperature records under “Forehead mode” 

 

°C / °F Switch mode 

 

Battery: one lithium cell battery (CR2032 *1pcs) 

 

This thermometer is an adjusted mode thermometer that converts the forehead temperature to 
display its “oral equivalent.” (according to the result of the clinical evaluation to get the offset value) 

 

Enclosure Rating: IP22 

 

Dimensions: 104.4 x 68.9 x 28.1 mm 

 

Weight: 57.4 grams including battery 

 

Battery life: around 3,000 continuous readings.   

 

Expected Service Life: 4 years 

 

The device should not submerge into any liquids and expose it to direct moisture.

 

 

There is no gender and age limitation for using infrared thermometer.   

  This is not an AP

 

or APG product.

 

   

 

Figure 1       Figure2        