background image

0,1 

N/A 

0,38 

0,73 

N/A 

1,2 

2,3 

10 

N/A 

3,8 

7,3 

100 

N/A 

12 

23 

For transmitters rated at a maximum output power not listed above, the recommended separation distance 

in meters (m) can be estimated  using the equation 

applicable to the frequency of the transmitter, where 

is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.   

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. 
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

 

 

Manufacturer’s declaration-electromagnetic immunity 

Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment 

The TMS series is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of the TMS series 
should assure that it is used in such an environment.

 

Test frequency 

(MHz) 

Band 

a)

 

(MHz) 

Service 

a)

 

Modulation 

b)

 

Maximum power 

(W) 

Distance 

(m) 

IMMUNITY TEST 

LEVEL 

(V/m) 

Compliance LEVEL 

(V/m) 

(for home healthcare) 

385 

380

–390 

TETRA 400 

Pulse 

modulation b) 

18 Hz 

1,8 

0,3 

27 

27 

450 

430

–470 

GMRS 460, 

FRS 460 

FM c) 

 

1 kHz sine 

0,3 

28 

28 

710 

704

–787 

LTE Band 13,17 

Pulse 

modulation b) 

217 Hz 

0,2 

0,3 

745 

780 

810 

800

–960 

GSM 800/900, 

TETRA 800, 

iDEN 820, 

CDMA 850, 

LTE Band 5 

Pulse 

modulation b) 

18 Hz 

0,3 

28 

28 

870 

930 

1720 

1700

– 

1990 

GSM 1800; 

CDMA 1900; 

GSM 1900;   

DECT; 

LTE Band 1, 3, 

4, 25; UMTS 

Pulse 

modulation b) 

217 Hz 

0,3 

28 

28 

1845 

1970 

2450 

2400

– 

2570 

Bluetooth,   

WLAN, 

802.11 b/g/n,   

RFID 2450,   

LTE Band 7 

Pulse 

modulation b) 

217 Hz 

0,3 

28 

28 

5240 

5100

– 

5800 

WLAN 802.11 

a/n 

Pulse 

modulation b) 

217 Hz 

0,2 

0,3 

5500 

5785 

NOTE    If  necessary  to  achieve  the  IMMUNITY  TEST  LEVEL, the  distance  between  the  transmitting  antenna  and  the ME  EQUIPMENT  or ME 
SYSTEM  may be reduced to 1 m.  The 1 m  test  distance is  permitted by IEC 61000-4-3. 

a) For some services, only the uplink frequencies are included. 
b) The carrier shall be modulated using a 50 % duty cycle square wave signal. 

c) As  an  alternative to FM modulation,  50 %  pulse modulation  at  18 Hz may  be  used  because  while  it  does  not  represent actual modulation,   

it  would  be  worst  case. 

 

Reviews: