Symbol Descriptions 

 

The CE mark and Notified Body 
Registration Numbers, the 
requirement of Annex II from Medical 
Device Directive 93/42/EEC are met.

 

 

Indicates this device is subject to the Waste 
Electrical and Electronic 
Equipment Directive in the European Union. 
To protect the environment, dispose of 
useless device at appropriate collection sites 
according to national or local regulations. 

 

Authorized 
representative in the 
European community 

 

Caution 

 

Please read the instructions for use 

 

Manufacturer 

 

Battery Recycling 

 

Paper Recycling 

IP22

 

Classification for water 
ingress and particulate 
matter. 

 

Manufacturer’s declaration-electromagnetic emissions 

The TMS series is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of the TMS series 
should assure that it is used in such an environment. 

Emissions test 

Compliance

 

Electromagnetic environment 

– 

guidance 

(for home healthcare environment)

 

RF emissions 
CISPR 11 

Group 1 

The TMS series uses RF energy only for its internal function. Therefore, its RF emissions are 
very low and are not likely to cause any interference in nearby electronic equipment.   

RF emissions 
CISPR 11 

Class B 

The TMS series is suitable for use in all establishments, including domestic establishments and 
those directly connected to the public low-voltage power supply network that supplies buildings 
used for domestic purposes. 

 

Manufacturer’s declaration – electromagnetic immunity 

The TMS series is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of the TMS series 
should assure that it is used in such an environment. 

Immunity test 

IEC 60601 test level

 

Compliance level

 

Electromagnetic environment 

– 

guidance 

(for home healthcare environment)

 

Electrostatic discharge(ESD) 

IEC 61000-4-2 

Contact:±8 kV   

Air±2 kV,±4 kV,±8 

kV,±15 kV 

Contact:±8 kV   

Air±2 kV,±4 kV,±8 

kV,±15 Kv 

Floors should be wood, concrete or ceramic tile. If floors are covered 

with synthetic material, the relative humidity should be at least 30% 

Power frequency(50, 60 Hz) 

magnetic field IEC 61000-4-8 

30 A/m 

50 Hz or 60 Hz 

30 A/m 

50 Hz and 60 Hz 

The TMS series power frequency magnetic fields should be at levels 

characteristic of a typical location in a typical home healthcare 

environment. 

NOTE UT is the a.c. mains voltage prior to application of the test level. 

 

Manufacturer’s declaration – electromagnetic immunity 

The TMS series is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of the TMS series 
should assure that it is used in such an environment. 

Immunity test 

IEC 60601 test level

 

Compliance 

level

 

Electromagnetic environment 

– 

guidance 

(for home healthcare environment)

 

Radiated RF   

IEC 61000-4-3 

10 V/m 

80 MHz 

– 2,7 GHz   

80 % AM at 1 kHz 

10 V/m 

80 MHz 

– 2,7 

GHz   

80 % AM at 1 

kHz   

Recommended separation distance: 

d = 1,2 

√P

 

d = 1,2 

√P

   

80MHz to 800 MHz 

d = 2,3 

√P

    800MHz to 2,7 GHz   

Where

 P

 is the maximum output power rating of the transmitter in watts (W) according to 

the transmitter manufacturer and 

is the recommended separation distance in metres 

(m). 
Field strengths from fixed RF transmitters, as determined by an electromagnetic site 
survey, 

a

 should be less than the compliance level in each frequency range.

Interference may occur in the vicinity of equipment marked with the 
following symbol: 

NOTE1    At 80 MHz and 800 MHz, the higher frequency range applies. 

NOTE2    These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects 

and people. 

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones  and land mobile radios, amateur radio, AM and FM 
radio  broadcast  and  TV  broadcast  cannot  be  predicted  theoretically  with  accuracy.  To  assess  the  electromagnetic  environment  due  to  fixed  RF 
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the TMS series is used exceeds 
the applicable RF compliance level above, the TMS series should be observed to verify normal operation. If abnormal performance is observed, additional 
measures may be necessary, such as re-orienting or relocating the TMS series. 

 

Recommended separation distances between 

portable and mobile RF communications equipment and the 

TMS series 

The TMS series is intended for use in an electromagnetic environment (for home healthcare) in which radiated RF disturbances are controlled. The 
customer or the user of the TMS series can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile 
RF communications equipment (transmitters) and the TMS series as recommended below, according to the maximum output power of the 
communications equipment. 

Rated maximum output power of 

transmitter 

Separation distance according to frequency of transmitter 

150 kHz to 80 MHz 

d = 

1,2 

 

80 MHz to 800 MHz 

d = 

1,2 

 

800 MHz to 2,7 GHz 

d = 

2,3 

 

0,01 

N/A 

0,12 

0,23 