background image

Battery replacement 

 

 

This device is supplied with one lithium cell CR2032x1. 

1

 Open the battery cover: Insert a pointed object, such as a pen, into battery cover pick hole.   

At the same time, use thumb to push battery cover out.   

2

 Hold the device and flip the battery out with a small screwdriver   

3

 Insert the new battery under the metal hook and press the other side of the battery down until you hear a click. 

4

 Replace the battery cover 

 

The positive (+) side up and the negative (-) side pointed down.

   

 

 

 

Trouble shooting: 

Error Message 

Problem 

Solution 

 

Error 5~9, the system is not functioning properly. 

Unload the battery, wait for 1 minute and repower it. If the message 

reappears, contact the retailer for service. 

 

Measurement before device stabilization.

 

Wait

 

for “Er1” to disappear.

 

 

The ambient temperature is not within the range between 10

o

C and 40

o

(50

o

F ~104

o

F). 

Allow the thermometer to rest in a room for at least 15 minutes at 

room temperature: 10

o

C and 40

o

C (50

o

F ~104

o

F). 

 

(1)

 

In ear/forehead mode: Temperature taken is higher than +42.2

o

(108°F) 

(2)

 

In surface mode: Temperature taken is higher than +80

o

C (176°F)  Please select the target within specifications. If a malfunction still 

exists, please contact the nearest retailer. 

 

(1)

 

In ear/forehead mode: Temperature taken is lower than +34

o

(93.2°F) 

(2)

 

In surface mode: Temperature taken is lower than –22

o

C (-7.6°F) 

 

Device cannot be powered on to the ready stage. 

Change with a new battery. 

 

Warranty:

 

Warranty: 12 months   

Manufacture Date: as the serial number (please open the battery cover, it is shown on the inside of the device.)

 

 

Ex.SN:E512A000001, the first “E” is External, the second number “5” is the last number of manufacture year, the third and the fourth number “12” is the manufacture 

month, the others is the serial number.

 

Note: The thermometer is calibrated at the time of manufacture. If you question calibration mode, the accuracy of temperature measurements or unexpected 

events at any time, please contact the dealers or nearest service address. 

Please report to the manufacturer and the competent authority of the Member State in which you are established about any serious incident that has 

occurred in relation to this device. 

 

 

Warning: No modification of this equipment is allowed.

 

 

Please read the instructions for use          BF type applied part 

IMPORTER 

Company Name?             

Company Name? 

General Information: Address? Website? 

 

 

 

 

Radiant Innovation Inc.     

Http://www.radiantek.com.tw 

Add: 1F, No.3, Industrial East 9

th

 Road,   

Science-Based Industrial Park, HsinChu, Taiwan 300. 

Medical Technology Promedt Consulting GmbH   

Add: Altenhofstrasse 80, D-66386 St. Ingbert, 

Germany 

 

Date of issue: 2020-06-23 

Ref No.: 062020 

Reviews: