Important Notes
Care and cleaning
T
he probe is the most delicate part of the thermometer. Use with care when cleaning the lens to avoid damage.
a. Please use alcohol swabs or cotton swabs moistened with 70%~75% alcohol to clean the probe lens.
b. Allow the probe to fully dry for at least 1 minute.
c. Storage temperature range: It should be stored at room temperature between -20~+50°C, RH
≤
85
%
d .Keep the unit dry and away from any liquids and direct sunlight.
e. The Probe should not be submerged into liquids.
* Note: Please check the device if damaged once it falls. If you can’t make sure of it ,please send the complete device
to the nearest retailer for recalibration.
Holding the thermometer too long may cause a higher ambient temperature reading of the probe.
This could make the body temperature measurement lower than usual.
Battery replacement
This device is supplied with one lithium cell CR2032x1.
○
1
Open the battery cover: Insert a pointed object, such as a pen, into battery cover pick hole.
At the same time, use thumb to push battery cover out.
○
2
Hold the device and flip the battery out with a small screwdriver
○
3
Insert the new battery under the metal hook and press the other side of the battery down until you hear
a click.
○
4
Replace the battery cover
The positive (+) side up and the negative (-) side pointed down.
Trouble shooting:
Error Message
Problem
Solution
Error 5~9, the system is not functioning properly.
Unload the battery, wait for 1 minute and repower it. If the message
reappears, contact the retailer for service.
Measurement before device stabilization.
Wait
for “Er1” to disappear.
The ambient temperature is not within the range between 10
o
C and 40
o
C (50
o
F ~104
o
F).
Allow the thermometer to rest in a room for at least 15 minutes at room
temperature: 10
o
C and 40
o
C (50
o
F ~104
o
F).
(1)
In ear/forehead mode: Temperature taken is higher than +42.2
o
C (108°F)
(2)
In surface mode: Temperature taken is higher than +80
o
C (176°F)
Please select the target within specifications. If a malfunction still
exists, please contact the nearest retailer.
(1)
In ear/forehead mode: Temperature taken is lower than +34
o
C (93.2°F)
(2)
In surface mode: Temperature taken is lower than –22
o
C (-7.6°F)
Device cannot be powered on to the ready stage.
Change with a new battery.
Warranty:
Warranty: 12 months
Manufacture Date: as the serial number (please open the battery cover, it is shown on the inside of the device.)
Ex.SN:E512A000001, the first “E” is External, the second number “5” is the last number of manufacture year, the third and the fourth number “12” is the manufacture
month, the others is the serial number.
Note: The thermometer is calibrated at the time of manufacture. If you question calibration mode, the accuracy of temperature measurements or unexpected
events at any time, please contact the dealers or nearest service address.
Please report to the manufacturer and the competent authority of the Member State in which you are established about any serious incident that has
occurred in relation to this device.
Warning: No modification of this equipment is allowed.
Please read the instructions for use BF type applied part
IMPORTER
Company Name?
Company Name?
General Information: Address? Website?
Radiant Innovation Inc.
Http://www.radiantek.com.tw
Add: 1F, No.3, Industrial East 9
th
Road,
Science-Based Industrial Park, HsinChu, Taiwan 300.
Medical Technology Promedt Consulting GmbH
Add: Altenhofstrasse 80, D-66386 St. Ingbert,
Germany
Date of issue: 2020-06-23