background image

Important Notes 

Care and cleaning

 

  T

he probe is the most delicate part of the thermometer. Use with care when cleaning the lens to avoid damage.   

  a. Please use alcohol swabs or cotton swabs moistened with 70%~75% alcohol to clean the probe lens. 

  b. Allow the probe to fully dry for at least 1 minute. 

  c. Storage temperature range: It should be stored at room temperature between -20~+50°C, RH

85

%

 

  d .Keep the unit dry and away from any liquids and direct sunlight.   

  e. The Probe should not be submerged into liquids. 

* Note: Please check the device if damaged once it falls. If you can’t make sure of it ,please send the complete device   

     to the nearest retailer for recalibration.   

  Holding the thermometer too long may cause a higher ambient temperature reading of the probe.   

  This could make the body temperature measurement lower than usual. 

Battery replacement 

 

 

This device is supplied with one lithium cell CR2032x1.

 

1

 Open the battery cover: Insert a pointed object, such as a pen, into battery cover pick hole.   

   At the same time, use thumb to push battery cover out.   

2

 Hold the device and flip the battery out with a small screwdriver   

3

 Insert the new battery under the metal hook and press the other side of the battery down until you hear     

   a click. 

4

 Replace the battery cover 

  The positive (+) side up and the negative (-) side pointed down.

   

 

 

 

 

Trouble shooting: 

Error Message 

Problem 

Solution 

 

Error 5~9, the system is not functioning properly. 

Unload the battery, wait for 1 minute and repower it. If the message 

reappears, contact the retailer for service. 

 

Measurement before device stabilization.

 

Wait

 

for “Er1” to disappear.

 

 

The ambient temperature is not within the range between 10

o

C and 40

o

C (50

o

F ~104

o

F). 

Allow the thermometer to rest in a room for at least 15 minutes at room 

temperature: 10

o

C and 40

o

C (50

o

F ~104

o

F). 

 

(1)

 

In ear/forehead mode: Temperature taken is higher than +42.2

o

C (108°F) 

(2)

 

In surface mode: Temperature taken is higher than +80

o

C (176°F) 

Please select the target within specifications. If a malfunction still 

exists, please contact the nearest retailer. 

 

(1)

 

In ear/forehead mode: Temperature taken is lower than +34

o

C (93.2°F) 

(2)

 

In surface mode: Temperature taken is lower than –22

o

C (-7.6°F) 

 

Device cannot be powered on to the ready stage. 

Change with a new battery. 

 

Warranty:

 

Warranty: 12 months   

Manufacture Date: as the serial number (please open the battery cover, it is shown on the inside of the device.)

 

 

Ex.SN:E512A000001, the first “E” is External, the second number “5” is the last number of manufacture year, the third and the fourth number “12” is the manufacture 

month, the others is the serial number.

 

Note: The thermometer is calibrated at the time of manufacture. If you question calibration mode, the accuracy of temperature measurements or unexpected 

events at any time, please contact the dealers or nearest service address. 

Please report to the manufacturer and the competent authority of the Member State in which you are established about any serious incident that has 

occurred in relation to this device. 

 

 

Warning: No modification of this equipment is allowed.

 

 

Please read the instructions for use          BF type applied part 

IMPORTER 

Company Name?             

Company Name? 

General Information: Address? Website? 

 

Radiant Innovation Inc.     

Http://www.radiantek.com.tw 

Add: 1F, No.3, Industrial East 9

th

 Road,   

Science-Based Industrial Park, HsinChu, Taiwan 300. 

Medical Technology Promedt Consulting GmbH   

Add: Altenhofstrasse 80, D-66386 St. Ingbert, 

Germany 

Date of issue: 2020-06-23 

Ref No.: 062020 

Summary of Contents for TH50Z

Page 1: ...n s condition seek medical assistance YELLOW Caution Prompt response by the OPERATOR is required GREEN Ready for use Memory Locations There are total 9 sets memories for body temperature Each memory also records the measurement date time mode icon and site location à When in power on press the ON MEM button to see the temperature stored with icon C F Switch à In power off mode The LCD only shows t...

Page 2: ...l aiming towards the membrane of the eardrum to obtain an accurate reading Fig 4 1 2 4 1 2 Press and hold the Ear button until you hear a beep sound Remove the probe from ear and read the temperature measurement on the LCD 4 1 3 After two short beeps are heard the icon stops flashing to be ready for next measurement 4 2 Forehead temperature measuring points for attention Remarks a Forehead tempera...

Page 3: ...t within the range between 10o C and 40o C 50o F 104o F Allow the thermometer to rest in a room for at least 15 minutes at room temperature 10o C and 40o C 50o F 104o F 1 In ear forehead mode Temperature taken is higher than 42 2o C 108 F 2 In surface mode Temperature taken is higher than 80o C 176 F Please select the target within specifications If a malfunction still exists please contact the ne...

Page 4: ...he TH5xy series should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance for home healthcare environment Electrostatic discharge ESD IEC 61000 4 2 Contact 8 kV Air 2 kV 4 kV 8 kV 15 kV Contact 8 kV Air 2 kV 4 kV 8 kV 15 kV Floors should be wood concrete or ceramic tile If floors are covered with synthetic material...

Page 5: ... transmitter in watts W according to the transmitter manufacturer NOTE1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people Manufacturer s declaration electromagnetic immunity Test specifications for ENCLOSUR...

Reviews: