Important Notes 

Cleaning 
And Storage

 

Please make sure the probe is clean to ensure an accurate reading. 

 

The probe lens is the most delicate part of the thermometer. Use with care when cleaning the probe lens to avoid damage. 

a. Please use alcohol swabs or cotton swabs moistened with 70%~75% alcohol to clean the probe lens. 
b. Allow the probe to fully dry for at least 1 minute. 
c. Keep the unit dry and away from any liquids and direct sunlight.   

        d.

 

Storage temperature range: It should be stored at room temperature between -20~+50°C,

 

RH

85% 

e. The Probe should not be submerged into liquids. 

  Holding the thermometer too long may cause a higher ambient temperature reading of the probe. 

        This could make the body temperature measurement lower than usual.

 

Battery replacement 
 

 

This device is supplied with one lithium cell CR2032x1. 
1. Open the battery cover: Use the thumbs to push battery cover out. (See Figure 1)   
2. Hold the device and flip the battery out with a small screwdriver. (See Figure 2) 
3. Insert a new battery under the metal hook and press down until you hear a “click” sound. (See Figure 3) 
4. Replace the battery cover. 

         

         

 

      Figure 1                        Figure 2                        Figure 3 

 

 

The positive (+) side up and the negative (-) side pointed down.

 

 

Trouble shooting: 

Error Message 

Problem 

Solution 

 

Error 5~9, the system is not functioning properly. 

Unload the battery, wait for 1 minute and repower it. If the message 
reappears, contact the retailer for service. 

 

Measurement before device stabilization.

 

Wait

 

for “Er1” to disappear.

 

 

The ambient temperature is not within the range between 10

o

C and 40

o

C. 

Allow the thermometer to rest in a room for at least 15 minutes at room 
temperature: 10

o

C and 40

o

C. 

 

Temperature taken is higher than +43

o

Please select the target within specifications. If a malfunction still exists, 
please contact the nearest retailer. 

 

Temperature taken is lower than +28

o

 

Device cannot be powered on to the ready stage. 

Change with a new battery. 

 

Warranty:

 

Warranty: 12 months   
Manufacture Date: as the serial number (please open the battery cover, it is shown on the inside of the device.)

 

 

Ex.SN:E512A000001, the first “E” is External, the second number “5” is the last number of manufacture year, the third and the fourth number “12” is the manufacture 
month, the others is the serial number.

 

Note: The thermometer is calibrated at the time of manufacture. If you question calibration mode, the accuracy of temperature measurements or unexpected   
          events at any time

,

 please contact the dealers or nearest service address.

   

Please report to the manufacturer and the competent authority of the Member State in which you are established about any serious incident that has   
occurred in relation to this device. 

 

 

Warning: No modification of this equipment is allowed. 

 

Follow instructions for use           

  BF type applied part 

 

 

IMPORTER 
Company Name?                         

Company Name? 
General Information: Address? Website? 

Radiant Innovation Inc.     
Http://www.radiantek.com.tw 
Add: 1F, No.3, Industrial East 9

th

 Road,   

Science-Based Industrial Park, HsinChu, Taiwan 300. 

Medical Technology Promedt Consulting GmbH   
Add: Altenhofstrasse 80, D-66386 St. Ingbert, 
Germany 

 

 

 

Date of issue: 2021-07-14 
Ref No.: 072021 

Summary of Contents for TH09F

Page 1: ...m from the central forehead Fig 1 and press the START button to get the forehead measurement The time consuming for measurement might be 1 second After two short beeps are heard the icon stops flashing to be ready for next measurement Figure 1 Points for attention a Forehead temperature is displayed in oral mode This mode converts the forehead temperature to display its oral equivalent value b Bef...

Page 2: ...it for Er1 to disappear The ambient temperature is not within the range between 10o C and 40o C Allow the thermometer to rest in a room for at least 15 minutes at room temperature 10o C and 40o C Temperature taken is higher than 43o C Please select the target within specifications If a malfunction still exists please contact the nearest retailer Temperature taken is lower than 28o C Device cannot ...

Page 3: ...ic tile If floors are covered with synthetic material the relative humidity should be at least 30 Power frequency 50 60 Hz magnetic field IEC 61000 4 8 30 A m 50 Hz or 60 Hz 30 A m 50 Hz and 60 Hz The TH09F power frequency magnetic fields should be at levels characteristic of a typical location in a typical home healthcare environment Manufacturer s declaration electromagnetic immunity The TH09F i...

Page 4: ...er s declaration electromagnetic immunity Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment The TH09F is intended for use in the electromagnetic environment for home healthcare specified below The customer or the user of the TH09F should assure that it is used in such an environment Test frequency MHz Band a MHz Service a Modulation b Maximum power W Distance ...

Reviews: