Recommended separation distances between 

portable and mobile RF communications equipment and the TH09F

 

The TH09F is intended for use in an electromagnetic environment (for home healthcare) in which radiated RF disturbances are controlled. The customer 
or the user of the TH09F can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF 
communications equipment (transmitters) and the TH09F as recommended below, according to the maximum output power of the communications 
equipment. 

Rated maximum output power of 

transmitter 

Separation distance according to frequency of transmitter 

150 kHz to 80 MHz 

d = 

1,2 

 

80 MHz to 800 MHz 

d = 

1,2 

 

800 MHz to 2,7 GHz 

d = 

2,3 

 

0,01 

N/A 

0,12 

0,23 

0,1 

N/A 

0,38 

0,73 

N/A 

1,2 

2,3 

10 

N/A 

3,8 

7,3 

100 

N/A 

12 

23 

For transmitters rated at a maximum output power not listed above, the recommended separation distance 

in meters (m) can be estimated  using the equation 

applicable to the frequency of the transmitter, where 

is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.   

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. 
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

 

 

Manufacturer’s declaration-electromagnetic immunity 

Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment 

The TH09F is intended for use in the electromagnetic environment (for home healthcare) specified below.The customer or the user of the TH09F should 
assure that it is used in such an environment.

 

Test frequency 

(MHz) 

Band 

a)

 

(MHz) 

Service 

a)

 

Modulation 

b)

 

Maximum power 

(W) 

Distance 

(m) 

IMMUNITY TEST 

LEVEL 

(V/m) 

Compliance LEVEL 

(V/m) 

(for home healthcare) 

385 

380

–390 

TETRA 400 

Pulse 

modulation b) 

18 Hz 

1,8 

0,3 

27 

27 

450 

430

–470 

GMRS 460, 

FRS 460 

FM c) 

 

1 kHz sine 

0,3 

28 

28 

710 

704

–787 

LTE Band 13,17 

Pulse 

modulation b) 

217 Hz 

0,2 

0,3 

745 

780 

810 

800

–960 

GSM 800/900, 

TETRA 800, 

iDEN 820, 

CDMA 850, 

LTE Band 5 

Pulse 

modulation b) 

18 Hz 

0,3 

28 

28 

870 

930 

1720 

1700

– 

1990 

GSM 1800; 

CDMA 1900; 

GSM 1900;   

DECT; 

LTE Band 1, 3, 

4, 25; UMTS 

Pulse 

modulation b) 

217 Hz 

0,3 

28 

28 

1845 

1970 

2450 

2400

– 

2570 

Bluetooth,   

WLAN, 

802.11 b/g/n,   

RFID 2450,   

LTE Band 7 

Pulse 

modulation b) 

217 Hz 

0,3 

28 

28 

5240 

5100

– 

5800 

WLAN 802.11 

a/n 

Pulse 

modulation b) 

217 Hz 

0,2 

0,3 

5500 

5785 

NOTE    If  necessary  to  achieve  the  IMMUNITY  TEST  LEVEL, the  distance  between  the  transmitting  antenna  and  the ME  EQUIPMENT  or ME 
SYSTEM  may be reduced to 1 m.  The 1 m  test  distance is  permitted by IEC 61000-4-3. 

a) For some services, only the uplink frequencies are included. 
b) The carrier shall be modulated using a 50 % duty cycle square wave signal. 

c) As  an  alternative to FM modulation,  50 %  pulse modulation  at  18 Hz may  be  used  because  while  it  does  not  represent actual modulation,   

it  would  be  worst  case. 

 

Summary of Contents for TH09F

Page 1: ...m from the central forehead Fig 1 and press the START button to get the forehead measurement The time consuming for measurement might be 1 second After two short beeps are heard the icon stops flashing to be ready for next measurement Figure 1 Points for attention a Forehead temperature is displayed in oral mode This mode converts the forehead temperature to display its oral equivalent value b Bef...

Page 2: ...it for Er1 to disappear The ambient temperature is not within the range between 10o C and 40o C Allow the thermometer to rest in a room for at least 15 minutes at room temperature 10o C and 40o C Temperature taken is higher than 43o C Please select the target within specifications If a malfunction still exists please contact the nearest retailer Temperature taken is lower than 28o C Device cannot ...

Page 3: ...ic tile If floors are covered with synthetic material the relative humidity should be at least 30 Power frequency 50 60 Hz magnetic field IEC 61000 4 8 30 A m 50 Hz or 60 Hz 30 A m 50 Hz and 60 Hz The TH09F power frequency magnetic fields should be at levels characteristic of a typical location in a typical home healthcare environment Manufacturer s declaration electromagnetic immunity The TH09F i...

Page 4: ...er s declaration electromagnetic immunity Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment The TH09F is intended for use in the electromagnetic environment for home healthcare specified below The customer or the user of the TH09F should assure that it is used in such an environment Test frequency MHz Band a MHz Service a Modulation b Maximum power W Distance ...

Reviews: