9. Product Inspection
Symptom
Cause
Action
When Display
trembles.
Load cell is damaged.
Load cell insulation resistance.
Indirect occurrence
Load cell input, output.
Check resistance
Check load cell’s insulation resistance.
When weight goes
up at a regular ratio
or zero returns are
not made.
Loadcell faulty
Check load cell’s insulation resistance.
Loadcell connection is
insufficient.
Check the wiring between load cell and the
main device.
Check the load cell’s calbe’s disconnection.
When weight
changes into (-).
Loadcell wiring is
reversed.
Check load cell’s ouput cable connection.
Displayed as Lc_Er
Load cell is damaged.
Load cell connection is
bad.
Check the load cell’s condition and calbe
connection.
Displayed as c_Err
Actual calibration set
below zero (-)
Performing actual load calibration at plus(+)
If the number on
the Display does
not match
Out of maximum display
range(-19999 ~ 99999)
Re-check the 8-4 Digital Calibration setting
and the F-51 Force unit setting.
Restore to the settings saved by RESTORE execution
Pressing
key
in FUNC state does
not move to
calibration mode.
Locked State at 8-7
Lock Set-up.
State in which units of
conversion at F-51 Force
unit is established.
Unlock from Locking Settings at 8-7 Lock
Set-up.
Change to kgf unit (set value :0) at F-51.
Force unit
The number of
measurements is
incorrect
Maximum Display
Range Out of 99999
Check calibration settings and F-51
settings. (page 30~36), (page 29)
Restore to the settings saved by RESTORE
execution (page 37)
When the
measurement
changes are not
constant or
insignificant. (RI-
20W, 25W)
Sensor Cable Swapped
or Open
Sensor damaged or
faulty supply voltage
(EXC) occurred
Calibration setup error
• Check cabling status
• Checksensor supply voltage (EXC) and
output
• Check DIP S/W (page 9) (RI-20W)
• Check calibration settings (page 30-36)