RADIANQBIO INSTRUMENT www.radian21.com
25
F-03. Hold mode
(Standard setup value: 1)
Display data
Setting
0
Sample Hold :
To hold the display value at the time of Hold signal input.
1
Peak Hold (+) :
To hold the maximum value of display values during Hold signal input.
2
Absolute Peak Hold(+/-) :
To hold the maximum absolute value of display
values during Hold signal input.
3
A/D Peak Hold(+):
To hold the maximum A/D value during Hold signal input.
Hold.
Note) When set to A/D Peak Hold (+), the Hold operation is performed according to the A/D input (output
of the sensor), regardless of the sign of the F-50 Display reverse mode.
RI-10W, RI-20W, RI-25W, RI-27W, RI-148A,B : 200Hz
RI-50W
: 1kHz
F-04. BCD Busy time (BCD Output time setup)
(Standard setup value : 0.100)
Display data
Setting
0.050
50 ms
0.100
100 ms
0.200
200 ms
0.500
500 ms
1.000
1000 ms
F-10. Auto zero tracking (Auto zero operation range setup)
(Standard setup value : 00)
Display data
Setting
00
00
: Auto zero is not used
~
99
01 ~99 : Set up the operation range of auto zero
(If the indicator value is less than the set value, it
corresponds to the autozero operating range.)
F-11. Auto zero tracking time (Auto Zero tracking time setup)
(Standard setup value : 0.0)
Display data
Setting
0.0
0.0 ~5.0 : set up the auto zero operation time (0.0 ~ 5.0 sec)
(
When the indicator is in the autozero operating range, the display is
maintained by the operating time setting, and then autozero is
performed.)
~
5.0