7.3
Functionality
Item
Description
FFT ECG Analysis
The AED FFT ECG analysis system automatically
analyzes the patient’s ECG and advises you when
defibrillation is required
Waveform
Biphasic Truncated Exponential (BTE)
Impedance range
25
Ω
to 175
Ω
Electric shock energy
Adult mode: 150J
±
10%
Child mode: 50J
±
10%
Charging time
8 seconds from the initiation of charging to be available
for discharge (with a new, fully charged battery; with a
low-on battery, the charging time can be significantly
prolonged)
Operating time
For a new, fully charged battery pack at 20~ 30°C: 200
discharges at maximum energy or 10 hours of
continuous ECG monitoring.
Automated Self-tests
Monthly (every 1
st
day for each month):
Battery, charge/discharge, internal electronics
Audio Guidance/Alerts
Guidance for AED operation and CPR procedure
(voice record)
Beep sound alarm
Indicators
Overall status light (Battery, AED condition)
CPR procedure guidance LED
CPR procedure guidance drawings
ECG data record
13.75Mb (Max. 100 of ECG logs can be repeatedly
recorded)
Data communication
Bluetooth 2.0