Do not attempt to recharge or cause heat/damage the battery (BT-503)
To avoid possible fire or explosion hazard, do not use around the
following situation/place.
In the presence of flammable gases
In the presence of hyperbaric oxygen tank (within 10m)
In the presence of hydrogen tank (within 10m)
Do not use battery not provided by RADIAN
RADIAN is not responsible for damage/malfunction in
product operation caused by use of unofficial battery. Since use of
unofficial battery may cause trouble/malfunction, please use the
official battery provided by RADIAN
②
AED pads maintenance
AED pads’ expiration date shall be checked periodically. Expired
AED pads shall be replaced to new one immediately. Expiration date
is marked on the package
If sealing package of AED pads is ripped or damaged, do not use the
pads and replace it to new one. When AED pads are exposed to the
air, they can only be used within 24 hours
HR-503 only recognizes P-503 AED pads. Do not use any other AED
pads not provided by RADIAN
RADIAN recommends preparing additional pads with different
expiration dates for spare