Module Specification 1-19
module. If five additional modules are likewise loaded, then the
overall carrier dissipation is approximately 80W, which is within the
cooling available in most commercial VXIbus chassis. Apart from
the thermal issues, however, it is important to remember that a
fully enabled 1260-145F card would draw over 2.7A of current from
the +5V supply. If this is multiplied by six for the remaining slots, a
load of 16A would burden the +5V supply, exceeding the
specifications for the 1260-100 carrier. In reality, the worse-case
situation is not likely, since the 1260-145F is a matrix switch where
typically only one relay is engaged at a time and the power
dissipation falls considerably below that calculated in the example
above. For practical purposes, as the calculations can illustrate for
normal operation, the 1260-145F contributes a negligible amount
of thermal load on a chassis.
Introduction –
1260-145G
(1 – 12x12)
The 1260-145G is a plug-in matrix switch module developed for
the Racal Instruments 1260-100 Adapt-a-Switch Carrier. This
switch provides a 12x12 matrix with a maximum switching voltage
of 125VAC @ 0.3A or 30 VDC @ 1A.
The 1260-145G includes the following features:
•
Standard Adapt-a-Switch
plug-in design, providing for ease
of replacement
•
Data-Driven embedded descriptor, allowing immediate use with
any Option-01T switch controller, regardless of firmware
revision level.
Summary of Contents for 1260 VXI
Page 5: ......
Page 6: ...This page was left intentionally blank...
Page 10: ...1260 145 User Manual iv This page was left intentionally blank...
Page 34: ...1260 145 User Manual Module Specification 1 24 Racal Instruments 2003...
Page 60: ...1260 145 User Manual Optional Harness Assemblies 4 4 Racal Instruments 2003...