Publication No. 980673-055 Rev. A 1260-22 User Manual
EADS North America Test and Services Module Operation 3-1
Chapter 3
MODULE OPERATION
General
Information
The 1260-22 may be operated either in
message-based mode
or in
register-based mode
.
When the
message-based mode
of operation is used, commands
are sent to the 1260-01T command module. The 1260-01T
command module interprets the commands, and operates the
1260-22 module by sending 8-bit bytes to control registers on the
1260-22 module.
A conceptual view of the message-based mode of operation is
shown in
Figure 3-1
below.
Figure 3-1, Message-Based Mode of Operation
When
the
register-based mode
of operation is used, the user
writes to the control register on the 1260-22 module directly. The
1260-01T command module does not monitor the operations, and
does not track the state of the relays on the module in this mode.
A conceptual view of the register-based mode of operation is
shown in
Figure 3-2
below.
Figure 3-2, Register-Based Mode of Operation
Since the 1260-01T command module does not monitor the
register-based mode of operation, it is advisable to select
either
the message-based or the register-based mode of operation, and
continue to use the same mode throughout the application
program.
PC (MXI)
1260-01T
1260-22
VXIbus
"CLOSE (@7(1))"
Write value 2 to A24 Address 205001
PC (MXI)
1260-120
Write value 2 to A24 Address 205001