Page 4
06206-90(I)
EMC COMPLIANCE
This device complies with Part 15 of the FCC Rules. Operation
is subject to the following two conditions: (1) this device may not
cause harmful interference, and (2) this device must accept any
interference received, including interference that may cause
undesired operation.
This equipment has been tested and found to comply with the
limits for a Class A digital device, pursuant to part 15 of the FCC
Rules. These limits are designed to provide reasonable protection
against harmful interference when the equipment is operated in a
commercial environment. This equipment generates, uses, and
can radiate radio frequency energy and, if not installed and used
in accordance with the instruction manual, may cause harmful
interference to radio communications. Operation of this equipment
in a residential area is likely to cause harmful interference in which
case the user will be required to correct the interference at his own
expense.
This ISM device complies with Canadian /CES-001.
Cet apparei/ISM est conforme a Ia norme NMB-001 du Canada.
EN55011/CISPR 11, Group 1, Class B device
Class B equipment is suitable for use in domestic establishments
and in establishments directly connected to a low voltage power
supply network which supplies buildings used for domestic
purposes.
Note:
Instrument measurements may be disrupted by conducted
RF interference at 13, 35, and 54 MHz.
To meet EMC Compliance, a ferrite core (YOUNG 18500) must be
installed on the wall adapter DC power cable. The cable must pass
through the center hole of the choke at least 4 times, creating 3
loops around the outside as shown below.
18500 FERRITE CHOKE
WARRANTY
This product is warranted to be free of defects in materials and
construction for a period of 12 months from date of initial purchase.
Liability is limited to repair or replacement of defective item. A copy
of the warranty policy may be obtained from R. M. Young Company.
CE COMPLIANCE
This product has been tested and complies with European CE
requirements for the EMC Directive. Please note that shielded cable
must be used.