R-Can Sterilight SC1 Installation, Operation And Maintenance Manual Download Page 10

7

Quartz Sleeve Replacement / Cleaning:

1. If the water contains any hardness minerals  

 

    (calcium or magnesium), iron or manganese,  

 

    the quartz sleeve will require periodic 

  cleaning. To remove the quartz sleeve, first 

  remove the UV lamp as outlined in step 1-4  

 

    (page 6) then perform the following steps:

    Caution:

 prior to performing any work on the    

  disinfection system, always disconnect the power  

  supply first. 

  a) Shut off water supply and drain all lines.

    b) Remove aluminum gland nuts from chamber 

        (Figure 6), checking for the free floating spring inside 

        sleeve at the opposite end to the lamp connection 

        (do not allow quartz sleeve to fall).

    c) Carefully remove o-rings from the quartz sleeve 

        (Figure 7). As the o-ring may tend to adhere to the 

        quartz sleeve, it is recommended to replace the o-rings 

     annually. Remove quartz sleeve carefully from chamber.

    d) Clean the outside of the quartz sleeve with a cloth 

         soaked in CLR™ or Limeaway™. Repeat the process as 

         often as necessary to keep the quartz sleeve clean. Be 

         sure to remove all traces of cleaning fluid from the 

       sleeve before it is reinstalled in the reactor (be sure not 

        to allow liquid inside the sleeve).

    e) Re-assemble the quartz sleeve with spring in the UV  

      chamber allowing the  sleeve to protrude an equal 

      distance from both ends of the UV chamber (Figure 8).

    f) Wet the o-rings and slide onto each end of the quartz  

    sleeve and reassemble the gland nuts (hand tight is 

      sufficient).

    g) Re-tighten all connections, turn on water slowly and  

     check for leaks.

    h) Re-install the UV lamp and lamp connector as per prior 

         instructions.

    i) Plug in ballast and verify the Lamp/Power LED is 

       illuminated.

Note:

 

If the system is put on a temporary by-pass or if 

it becomes contaminated after the disinfection system, 

it will be necessary to shock the system with household 

bleach for a full 20 minutes before resuming the use of 

the water.

Figure 6

Figure 7

Figure 8

Figure 5

Summary of Contents for Sterilight SC1

Page 1: ...and Maintenance sc1 sc2 5 sc4 sc1 2 sc2 5 2 sc4 2 425 Clair Road West P O Box 1719 Guelph ON N1L 1R1 Canada t 519 763 1032 f 519 763 5069 e water r can com i www r can com Manufactured in Canada by EP...

Page 2: ...ronmentally Friendly lamps throughout our entire product offering Low mercury Hg technology provides lamps with less than 10 mg of mercury including our amalgam lamps up to 70 less mercury than the co...

Page 3: ...ns 2 Water Chemistry 3 Installing Your UV Disinfection System 4 5 Operating and Maintenance Instructions 6 7 Fluence Dose Curves 8 Manufacture s Warranty 9 Specifications 10 11 Symbols Caution Protect...

Page 4: ...or sc2 5 sc2 5 2 qs 330d for sc4 sc4 2 aluminum gland nut rn 001 304 stainless steel reactors with 600 grit polish A249 pressure rated tube 1 2 MNPT 3 8 FNPT combination ports for application flexibil...

Page 5: ...to disconnect 7 Do not use this disinfection system for other than intended use potable water applications The use of attachments not recommended or sold by the manufacturer distributor may cause an...

Page 6: ...ess is less than 7 gpg the UV unit should operate efficiently provided the quartz sleeve is cleaned periodically If total hardness exceeds 7 gpg the water should be softened If your water chemistry co...

Page 7: ...s the disinfection system should be connected to a ground fault interrupt circuit The disinfection system is intended for indoor use only do not install disinfection system where it may be exposed to...

Page 8: ...TTACHED TO THE DISINFECTION SYSTEM AS THIS COULD DAMAGE THE O RING SEALS 4 Prior to connecting the power source check all connections to ensure that they are indeed secure slowly turn on water supply...

Page 9: ...t to angle the lamp as it is removed from the chamber If the lamp is removed on an angle pressure will be applied on the inside of the quartz sleeve causing the sleeve to fracture 3 To install a new l...

Page 10: ...he outside of the quartz sleeve with a cloth soaked in CLR or Limeaway Repeat the process as often as necessary to keep the quartz sleeve clean Be sure to remove all traces of cleaning fluid from the...

Page 11: ...8 Fluence Dose Curves SC1 SC1 2 SC2 5 SC2 5 2 SC4 SC4 2...

Page 12: ...1 mg L f UV Transmittance 75 call factory for recommendations on applications where UVT 75 Where total hardness is less than 7 gpg the UV unit should operate efficiently provided the quartz sleeve is...

Page 13: ...1 8 kg 4 lbs 2 0 kg 4 5 lbs 2 9 kg 6 5 lbs Electrical Voltage 100 130V 50 60Hz 100 130V 50 60Hz 100 130V 50 60Hz Power Consumption 12 w 16 W 19 w Lamp Watts 10 W 14 w 17 w Maximum Operating Pressure...

Page 14: ...ght 1 8 kg 4 lbs 2 0 kg 4 5 lbs 2 9 kg 6 5 lbs Electrical Voltage 200 250V 50 60Hz 200 250V 50 60Hz 200 250V 50 60Hz Power Consumption 12 w 16 W 19 w Lamp Watts 10 W 14 w 17 w Maximum Operating Pressu...

Reviews: