User’s Guide
Normal use
This examination couch enables the doctor to carry out general medical treatment, and
meets the following important requirements :
Easy positioning of the patient
Close position of the practitioner
Only nursing staff, hospital service staff and technical service staff should handle or use
the examination couch.
Patients or people accompanying them must not operate the examination couch. They
must not touch the control switches or levers.
This equipment must not be used in the presence of an ignitable anaesthetic air mix,
or with oxygen or nitrous oxide.
Electrical connections
This equipment must be connected to the main electricity supply.
Connection to the supply :
120 V or 230V
(depending on country)
Frequency :
50/0Hz
Protection Classification / Degree of electric shock protection : Class 2 / B
Power absorbed :
300W
Intermittent function mode :
1min / min
Maximum patient weight of
200 kg
(tested at a static weight of
00 kg and tested at a dynamic weight of 250 kg).
User’s Guide – Examination Couch
Features
Electric variable height from 4cm to 2cm
Steel frame with light grey epoxy finish
Gas-spring assisted or electric backrest
Flat upholstery with class M1 flame retardant covering
Hand control
Adjustable levellers under base
Block’n Roll
®
-
Integrated retractable castor system
(ref: 2050-30 and 2050-50)
Paper roll holder (max length 50cm - ø 12.5)
Mobile hoist compatibility – clearance of 12cm
If the couch is not to be used for a prolonged period of time, it should be switched off and discon-
nected from the mains.