
Operation
34
►
Cut off the assembly shaft at the marked position on the valve (
/
QF5C, item 6.2 Figure 9).
►
Remove the screws (1 / PQ5U, item 2 Figure 10).
►
Remove the connecting nozzles (2 / PQ5U, item 1 Figure 10).
►
Replace the O-rings (3 / QF5C, item 6.5 Figure 9).
►
Mount the connecting nozzles (2 / PQ5U, item 1 Figure 10).
►
Fasten the screws (1 / PQ5U, item 2 Figure 9).
►
Mount the O-ring (1 / QF5C, item 6.4 Figure 9).
►
Mount the profile O-ring (2 / QF5C, item 6.3 Figure 9).
►
Mount the valve plate together with the valves (3 / QF5C, item 2; 6.4