Warning: the Control Unit should always be operating to prevent pressure
wounds.
In operate mode, press operate/standby button
for the system to enter
standby mode. The system should be in standby mode before shut down.
Switch the power switch
to off and the warning code“
”will appear
on the display to shut off the system.
Note: Power failure alarm will be triggered if the power is switched off in
operating mode (refer to
5.2.3 Audiovisual Alarm
). Press power switch
to
restart the system, or press Alarm Mute
to turn off the system (refer
5.2.4
Alarm Mute
).
Table 1 Weight and Comfort Level Reference
5.2.2
CPR
When CPR needs to be performed, quickly rotate the CPR valve to “OPEN”
position, at the same time, disconnect the hose connector from the Control Unit
to speed up the air release.
5.2.3
Audiovisual Alarm
Power Failure – When electrical shortages occur or power cord is unplugged
without turning off the Control Unit or
is pressed (intentionally or
unintentionally), the “Power Failure” indicator will light up along with buzzer
and will last 20 minutes.
NOTE: When the Control Unit has not been used for more than 3 months, the
Control Unit may need 6
hours or more of operating time for the alarms to function
properly.