Press this button to start operating or go into standby.
5.1.4
Auto-Firm
The Control Unit will go into the inflation mode (LED lights flashing) every time
the operate mode is triggered. This insures the mattress is able to reach its
maximum operating pressure. Once the max pressure level is reached, the
Control Unit will automatically switch into the previous selected mode and
comfort level. User can also use this function as full mattress inflation during
patient sit-up or ingress/egress for better support.
5.1.5
Function Mode Switch
Alternate – the air cells of the mattress will be proportionally deflated to
reduce the surface pressure. The alternating cycle will continue at the
selected cycle time until another mode is selected.
Static – the mattress maintains at the selected pressure. The Control Unit will
automatically fall back to alternation mode after 20 minutes.
Constant Low Pressure (CLP) – the mattress maintains at a reduced selected
pressure in static mode.
5.1.6
Panel Lock-Out
Press the Lock-Out button to lock the panel. Should the panel remain untouched
for 30 seconds; the Lock-Out feature will lock the panel to prevent anyone from
accidentally changing settings without notice. To unlock, press the Lock-Out
button for 3 seconds.
5.1.7
Comfort Level
Comfort level controls the air pressure output level. Press Firm button and the
output pressure will increase and higher pressure output will support a higher
weight; for decreasing air pressure, vice versa. Refer to
Table 1 Weight and
Comfort Level Reference
for weight and comfort level suggestion.
5.1.8
Auto Detection